Fisher Scientific Digital MultiTube Vortexer
The Fisher Scientific Digital MultiTube Vortexer operates in a defined 3.6mm eccentric orbit with a capacity of up to 50 samples at once. Speed range is from 500 to 2500 RPM. This MultiTube vortexer offers a powerful motor that performs quietly and continuously.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Fisher Scientific Digital MultiTube Vortexer includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific Digital MultiTube Vortexer Specifications
|
Speed Range |
500 to 2500 RPM |
|
Orbit |
3.6mm |
|
Weight Capacity |
10 lbs (4.5 kg) |
|
Programmable Timer |
1 sec to 160 hrs |
|
Tube Capacity |
Up to 50 |
|
Dimensions (W x L x H) |
38.1 x 24.1 x 40.6 cm |
|
Electrical Requirements |
120V, 50/60Hz, 75W, 5A |
- Verify power supply and PCB output correct, stable voltages across all required circuits
- Confirm the correct platform head is installed according to specifications
- Inspect the display for dim or dead pixels, replacing any compromised sections to maintain clear visibility
- Confirm full functionality of all buttons and knobs, ensuring accurate input response
- Verify that touch, pulse, and continuous modes operate fully and as intended
- Inspect all internal wiring for breakages, tears, or wear, replacing any underperforming components to ensure reliability
- Disassemble and thoroughly clean the internal casing to remove any accumulated debris
- Check that the feet are intact and confirm the vortexer is properly balanced on a flat surface
- Conduct a comprehensive cosmetic inspection to ensure all exterior surfaces meet quality standards
Technician Inspection and Recertification Checklist: Vortexer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Verify power supply and PCB output stable, correct voltages across circuits | Completed | |
| Confirm correct platform head is installed according to specifications | Completed | |
| Inspect display for dim or dead pixels; replace compromised sections for clarity | Completed | |
| Confirm functionality of all buttons and knobs for accurate input response | Completed | |
| Verify full operation of touch, pulse, and continuous modes as intended | Completed | |
| Inspect internal wiring for damage; replace underperforming components for reliability | Completed | |
| Disassemble and clean internal casing to remove accumulated debris | Completed | |
| Check feet for integrity and confirm vortexer is balanced on flat surface | Completed | |
| Conduct comprehensive cosmetic inspection to ensure quality of exterior surfaces | Completed |