Speed Range
250 to 3,000 RPM (250 to 2,000 RPM with included block)
Temperature Range
Ambient +5°C to 100°C
Heating Rate
5°C/minute
- Large color touchscreen provides clear visibility and intuitive control
- Interchangeable blocks with automatic recognition simplify setup and operation
- Mixing and temperature control support precise processing of small samples
- Easy-to-use interface with clear icons for Stop, Play, Pause, Skip, and Quick Mix
- Home screen displays current step number and program name for quick reference
- Unit operating status is clearly indicated with bright on-screen text
- Parameters can be adjusted instantly for individual steps during a run
- Minimum and maximum settings are automatically determined by block type
- Multi-step programs allow up to five steps with automatic repeat capability
Interchangeable blocks with automatic recognition ensure appropriate parameter limits are applied, helping simplify setup and protect samples. Users can create, edit, and save multi-step programs with up to five steps, adjust parameters for individual steps, and automatically repeat sequences as needed. Programs can be saved with descriptive names, stored internally for quick recall, and protected using a parameter lock. With precise speed control and a wide temperature range, the Thermal Mixer II delivers flexible, easy-to-use performance for routine laboratory workflows.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Fisher Scientific Heat Only Thermal Mixer II w/ 1.5ml Block includes:
- > 24 x 1.5ml Block
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific Heat Only Thermal Mixer II w/ 1.5ml Block Specifications
| Speed Range | 250 to 3,000 RPM (250 to 2,000 RPM with included block) |
| Speed Increments | 50 RPM |
| Temperature Range | Ambient +5°C to 100°C |
| Temperature Uniformity | ± 0.3°C between 20°C and 40°C |
| Temperature Accuracy |
± 0.2°C between 20°C and 40°C
± 0.5°C < 20°C and > 40°C |
| Temperature Increments | ± 0.1°C |
| Heating Rate | 5°C/minute |
| Timer | 0 to 99 hours, 59 minutes, 59 seconds, or continuous |
| Dimensions (W x D x H) | 9 x 13 x 5.3 (23 x 33 x 13.5 cm) |
| Weight | 25 lb. (11.34 kg) |
| Electrical Requirements | 12VDC via 115/230 AC Adapter, 50/60Hz, 5A, 120W |
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm stability and accuracy of power supply and PCB voltage outputs | Completed | |
| Inspect display integrity, ensuring pixel clarity and replacing any degraded elements | Completed | |
| Verify functionality and responsiveness of all control buttons, switches, and knobs | Completed | |
| Inspect internal wiring and connections for continuity, replacing or securing degraded components | Completed | |
| Examine block holder platform for alignment, obstructions, or irregularities | Completed | |
| Inspect motor bearings for wear or misalignment, replacing if necessary | Completed | |
| Calibrate rotational speed using an NIST-traceable tachometer | Completed | |
| Calibrate temperature with an NIST-traceable probe, validating accuracy | Completed | |
| Ensure timer precision by validating against set parameters | Completed | |
| Conduct cosmetic inspection for quality standard compliance | Completed |