Scientific Industries Vortex Genie 2 Vortexer
Base Dimensions
165 x 122 x 165mm (6.5 x 4.8 x 6.5in)
Weight
4 Kg (8.8 lb)
Speed
600-3200 RPM
- Tough metal housing combined with just enough weight, provides a stable platform for all types of mixing.
- Variable Speed Control allows slow speed shaking action up to high speed vortexing.
- A broad range of attachments of various shapes, sizes, and materials allows for mixing of almost all common tubes or containers.
- A 3-position power switch operates with continuous action or "Touch On" control. With the true "Hands-Free" operation of continuous action, lab operators can greatly increase lab efficiency.
- Unmatched Reliability with many years of proven performance in laboratories all over the world.
Scientific Industries Vortex Genie 2 Vortexer features variable Manual Speed Controls, Touch Function and Automatic Function, 3? Platform Head. The Fisher Vortex Genie 2 G560 offers a speed range of 600 to 3200 rpm. All of our equipment is fully tested before being shipped.
The Scientific Industries Vortex Genie 2 Vortexer includes:
- > Rubber Dimpled Head
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Scientific Industries Vortex Genie 2 Vortexer Specifications
|
Environmental |
0C - 38C |
|
Electrical Requirements |
120 V, 60 Hz, 0.65 amps |
|
Weight |
4 kg, 8.8 lbs |
|
Base Dimensions (D x W x H) |
6.5 x 4.8 x 6.5 in |
|
Speed |
600-3200 RPM |
- Verify power supply and PCB output correct, stable voltages across all required circuits
- Confirm the correct platform head is installed according to specifications
- Inspect the display for dim or dead pixels, replacing any compromised sections to maintain clear visibility
- Confirm full functionality of all buttons and knobs, ensuring accurate input response
- Verify that touch, pulse, and continuous modes operate fully and as intended
- Inspect all internal wiring for breakages, tears, or wear, replacing any underperforming components to ensure reliability
- Disassemble and thoroughly clean the internal casing to remove any accumulated debris
- Check that the feet are intact and confirm the vortexer is properly balanced on a flat surface
- Conduct a comprehensive cosmetic inspection to ensure all exterior surfaces meet quality standards
Technician Inspection and Recertification Checklist: Vortexer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Verify power supply and PCB output stable, correct voltages across circuits | Completed | |
| Confirm correct platform head is installed according to specifications | Completed | |
| Inspect display for dim or dead pixels; replace compromised sections for clarity | Completed | |
| Confirm functionality of all buttons and knobs for accurate input response | Completed | |
| Verify full operation of touch, pulse, and continuous modes as intended | Completed | |
| Inspect internal wiring for damage; replace underperforming components for reliability | Completed | |
| Disassemble and clean internal casing to remove accumulated debris | Completed | |
| Check feet for integrity and confirm vortexer is balanced on flat surface | Completed | |
| Conduct comprehensive cosmetic inspection to ensure quality of exterior surfaces | Completed |


Click Here to View the Technical Specs for the Vortex Genie 2 G560