VWR Analog Vortexer, Cat. #: 89399-880
Speed Range
On Mode: 300 to 2,500 RPM (continuous duty accessories)
Touch Mode: 300 to 3,500 RPM (cup head attachment)
Orbit
3.5 mm (0.13 in.)
Controls
Analog
- Ideal for mixing a number of sample container types
- Provides reliable, uniform mixing for repeatable results
- Operates in either “Touch” mode for momentary activation or “On” mode for continuous mixing
- Simple analog display for streamlined user operation
- Compact design conserves valuable bench space
All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR Analog Vortexer, Cat. #: 89399-880 includes:
- > Cup Head Attachment
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR Analog Vortexer, Cat. #: 89399-880 Specifications
| Speed Range |
On Mode: 300 to 2,500 RPM (continuous duty accessories)
Touch Mode: 300 to 3,500 RPM (cup head attachment) |
| Orbit | 3.5 mm (0.13 in.) |
| Timer | N/A |
| Controls | Analog |
| Operating Modes | On and touch |
| Dimensions (L x W x H) | 5.4 x 10.5 x 4.5 in. (26.7 x 13.7 x 11.4 cm) |
| Weight | 12 lb. (5.4 kg) |
| Electrical Requirements | 120V, 50/60Hz, 0.25A, 30W |
- Verify power supply and PCB output correct, stable voltages across all required circuits
- Confirm the correct platform head is installed according to specifications
- Inspect the display for dim or dead pixels, replacing any compromised sections to maintain clear visibility
- Confirm full functionality of all buttons and knobs, ensuring accurate input response
- Verify that touch, pulse, and continuous modes operate fully and as intended
- Inspect all internal wiring for breakages, tears, or wear, replacing any underperforming components to ensure reliability
- Disassemble and thoroughly clean the internal casing to remove any accumulated debris
- Check that the feet are intact and confirm the vortexer is properly balanced on a flat surface
- Conduct a comprehensive cosmetic inspection to ensure all exterior surfaces meet quality standards
Technician Inspection and Recertification Checklist: Vortexer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Verify power supply and PCB output stable, correct voltages across circuits | Completed | |
| Confirm correct platform head is installed according to specifications | Completed | |
| Inspect display for dim or dead pixels; replace compromised sections for clarity | Completed | |
| Confirm functionality of all buttons and knobs for accurate input response | Completed | |
| Verify full operation of touch, pulse, and continuous modes as intended | Completed | |
| Inspect internal wiring for damage; replace underperforming components for reliability | Completed | |
| Disassemble and clean internal casing to remove accumulated debris | Completed | |
| Check feet for integrity and confirm vortexer is balanced on flat surface | Completed | |
| Conduct comprehensive cosmetic inspection to ensure quality of exterior surfaces | Completed |


Click Here to View the Technical Specs for the VWR Analog Vortexer, Cat. #: 89399-880