Agilent 1260 Infinity I HPLC G1365C MWD Detector
Wavelength Range
190 to 950 nm
Detection Type
1024-element photodiode array
Data Rate
Up to 80 Hz
- Designed for highest optical performance, GLP compliance, and easy maintenance
- RFID tags on all flow cells and UV lamps for enhanced traceability
- Built-in holmium oxide filter enables fast wavelength accuracy verification
- Built-in temperature controls improve baseline stability
- Higher sensitivity and selectivity with simultaneous detection of up to eight wavelengths
- Low detection limits achieved through low noise electronics and patented flow cell design
- Programmable slit from 1 to 16 nm allows for rapid optimization of sensitivity and linearity
- Nine analytical and preparative flow cells for extensive application flexibility
- Extensive diagnostics, error detection, and display capabilities with Instant Pilot controller and Lab Advisor software
The device's built-in holmium oxide filter allows for rapid verification of wavelength accuracy, while integrated temperature controls enhance baseline stability. Diagnostic signals are available for monitoring temperature and lamp voltage, ensuring consistent operation. The detector's high sensitivity and selectivity are achieved through simultaneous detection of up to eight compound-specific wavelengths, and its advanced design includes low noise front-end electronics and a patented flow cell that significantly reduces short-term noise.
With its programmable slit ranging from 1 to 16 nm, users can easily optimize sensitivity and linearity to suit various applications. The MWD offers a wide linear range for accurate quantification of primary compounds, by-products, and impurities. It also features nine analytical and preparative flow cells for maximum flexibility, complemented by extensive diagnostics, error detection, and display capabilities via the Instant Pilot controller and Agilent Lab Advisor software, making the Agilent 1260 Infinity I HPLC G1365C MWD Detector an exceptional choice for sophisticated analytical tasks.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Agilent 1260 Infinity I HPLC G1365C MWD Detector includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Flow Cell
- > Tungsten and Deuterium Lamps
- > Lab Advisor Calibration Report
- > Lan Cable / Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity I HPLC G1365C MWD Detector Specifications
|
Detection Type |
1024-element photodiode array |
| Light Source | Deuterium and tungsten lamps |
| Data Rate | Up to 80 Hz |
| Wavelength Range | 190 to 950 nm |
| Short Term Noise (ASTM) |
< ± 0.7 x 10 -5 AU at 254 nm and 750 nm |
| Drift | < 0.9 x 10 -3 AU/h at 254 nm |
| Linear Absorbance Range | > 2 AU (5%) at 265 nm |
| Wavelength Accuracy | ± 1 nm |
| Wavelength Bunching | 1 to 400 nm |
| Slit Width | 1, 2, 4, 8, 16 nm |
| Diode Width | < 1 nm |
| Flow Cell | Standard: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum |
| Time Programmable | Wavelength, polarity, peak width, lamp bandwidth, autobalance, wavelength range, threshold, spectra storage mode |
| Dimensions (H x W x D) | 5.5 x 13.5 x 17 in. (14 x 34.5 x 43.5 cm) |
| Weight | 26.5 lb. (11.5 kg) |
| Electrical Requirements | 100-240V, 50/60Hz, 160VA |


Click Here to View the Technical Specs for the Agilent 1260 Infinity I HPLC G1365C MWD Detector