Agilent 1260 Infinity II HPLC G7115A DAD Detector
Wavelength Range
190 to 950 nm
Detection Type
1024-element photodiode array
Data Rate
120 Hz
- Engineered for peak optical performance and GLP compliance
- Easy maintenance with front access for quick component replacement
- 120 Hz data acquisition rate suitable for ultra-fast LC applications
- Long-life deuterium and tungsten lamps covering 190 to 950 nm wavelength range
- RFID tags on flow cells and UV lamps for enhanced traceability
- Simultaneous detection of up to eight wavelengths without loss of sensitivity
- Programmable slit adjustable from 1 to 16 nm for optimal sensitivity and resolution
- Built-in holmium oxide filter for fast wavelength accuracy checks
- Temperature control features for stable baselines under varying conditions
- Diagnostic tools for monitoring temperature and lamp voltage
Users benefit from the detector’s ability to maintain sensitivity while simultaneously analyzing up to eight wavelengths. The programmable slit, adjustable from 1 to 16 nm, allows fine-tuning of sensitivity, linearity, and spectral resolution. Maintenance is streamlined through easily accessible front components for quick replacement. It also incorporates a built-in holmium oxide filter for rapid wavelength accuracy verification, and temperature controls for enhanced baseline stability. Additional features include diagnostic signals for monitoring temperature and lamp voltage, thus bolstering the detector's reliability in fast-paced analytical environments.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Agilent 1260 Infinity II HPLC G7115A DAD Detector includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Flow Cell
- > Tungsten and Deuterium Lamps
- > Lab Advisor Calibration Report
- > Lan Cable / Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity II HPLC G7115A DAD Detector Specifications
|
Detection Type |
1024-element photodiode array |
| Light Source | Deuterium and tungsten lamps |
| Number of Signals | 8 |
| Data Rate | 120 Hz |
| Wavelength Range | 190 - 950 nm |
| Short Term Noise (ASTM) Single and Multi-Wavelength |
< ± 0.7 x 10 -5 AU, at 254 and 750 nm |
| Drift | < 0.9 x 10 -3 AU/h at 254 nm |
| Linear Absorbance Range | > 2 AU (5%) at 265 nm |
| Wavelength Accuracy | ± 1 nm |
| Wavelength Bunching | 1 - 400 nm, programmable in 1 nm steps |
| Slit Width | 1, 2, 4, 8, 16 nm |
| Diode Width | ~ 1 nm |
| Flow Cell | Standard: 13 ul volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum |
| Time Programmable | Wavelength, polarity, peak width, lamp bandwidth, autobalance, wavelength range, threshold, spectra storage mode |
| Dimensions (H x W x D) | 5.5 x 15.6 x 17 in. (14 x 39.6 x 43.6 cm) |
| Weight | 26.5 lb. (12 kg) |
| Electrical Requirements | 100-240V, 50/60Hz, 110VA |


Click Here to View the Technical Specs for the Agilent 1260 Infinity II HPLC G7115A DAD Detector