Corning PC-610D Digital Stirrer, 230V
Speed Range
60 to 1,150 RPM
Plate Dimensions
10 x 10 in. (25.4 x 25.4 cm)
Plate Material
Ceramic
- Durable ceramic top plate provides excellent chemical resistance and easy cleaning
- Digital display ensures precise monitoring of stirring speed
- Microprocessor control maintains consistent and repeatable performance
- User-friendly dial control allows easy adjustment of operating parameters
- Power indicator light ensures clear visibility of operational status
All of our equipment is in excellent condition and fully tested before shipping.
The Corning PC-610D Digital Stirrer, 230V includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Corning PC-610D Digital Stirrer, 230V Specifications
| Stirrer Speed | 60 to 1,150 RPM |
| Plate Dimensions | 10 x 10 in. (25.4 x 25.4 cm) |
| Plate Material | Ceramic |
| Control | Digital |
| Weight | 11.5 lb. (5.2 kg) |
| Dimensions (W x D x H) | 11 x 15.4 x 4.6 in. (19.7 x 39 x 11.8 cm) |
| Electrical Requirements | 230V, 50Hz, 0.3A, 73W |
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and CO₂ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed |


Click Here to View the Technical Specs for the Corning PC 420D Digital Hot Plate Stirrer, 230V