EC-135 Electrophoresis Power Supply
The EC Apparatus Electrophoresis Power Supply EC-135. Specifications: Maximum Voltage 250V, Voltage Steps Low Range Setting: 20-130V (Increments of 10V), Voltage Steps High Range Setting: 140-250V (Increments of 10V) Max. Current 300mA @ 150-250V; 500mA below 150V. Display Modes Voltage or current. Electrical Requirements 120V/50Hz.
Our equipment will be fully tested to perform at factory specifications before being shipped.
The EC-135 Electrophoresis Power Supply includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
|
EC-135 Electrophoresis Power Supply
|
Max Voltage: 250V
|
|
Max Current: 500mA
|
|
|
Max Wattage: 250W |
- Verify internal power supply output for correct voltage in alignment with equipment specifications
- Inspect LED display for any dim or dead pixels, ensuring optimal readability
- Examine internal wiring and connections, checking for any rips, tears, or compromised insulation
- Ensure fan is clean and fully operational to maintain adequate cooling
- Test PCB test points for proper voltage output, confirming consistent circuit performance
- Validate functionality of all buttons and switches, ensuring responsive control
- Measure and confirm constant voltage, amplitude, and wattage using precise resistance values
- Cross-check displayed voltage with multimeter readings for accuracy
- Monitor for rapid resistance changes, software errors, and power fluctuations
- Inspect line cord for any signs of wear, cuts, or damage impacting safety
- Confirm cosmetic condition meets quality standards, free of any damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Electrophoresis Power Supplies
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify internal power supply output voltage | Completed | |
| Inspect LED display for dim or dead pixels | Completed | |
| Examine internal wiring for rips or tears | Completed | |
| Ensure fan is clean and fully operational | Completed | |
| Test PCB test points for voltage accuracy | Completed | |
| Validate functionality of buttons and switches | Completed | |
| Confirm constant voltage, amplitude, and wattage with resistance | Completed | |
| Cross-check displayed voltage with multimeter | Completed | |
| Monitor for rapid resistance changes and power fluctuations | Completed | |
| Inspect line cord for wear or damage | Completed | |
| Confirm cosmetic condition meets quality standards | Completed |