Eppendorf 5301 Vacufuge Concentrator System
Maximum Speed
1,400 RPM
Heating Levels
Ambient, 30°C, 45°C, 60°C
Vacuum Level
< 20 mbar
- Specifically designed for gentle concentration of DNA, RNA, nucleotides, proteins, and gel electrophoresis samples
- Combines centrifugal force with vacuum to maximize efficiency in sample processing
- Features three aeration modes including aqueous, alcohol, and off for optimized evaporation
- Capable of reducing processing time by up to 20% to enhance overall laboratory workflow
- Offers a broad time range setting from 1 minute to 9 hours and 59 minutes plus a continuous mode for ultimate flexibility
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf 5301 Vacufuge Concentrator System includes:
- > F-45-48-11 Fixed Angle Rotor
- > Built in Pump
- > Operator's Manual
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5301 Vacufuge Concentrator System Specifications
| Maximum Speed | 1,400 RPM |
| Temperature Setpoints | Ambient, 30°C, 45°C, 60°C |
| Timer | 1 minute to 9 hours 59 minutes, or continuous |
| Ultimate Vacuum | < 20 mbar |
| Rotor Capacity | 48 x 1.5/2.0ml |
| Electrical Requirements | 115V, 60Hz, 370W |
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display panel for dim or dead pixels | Completed | |
| Confirm functionality of all buttons | Completed | |
| Check acrylic/glass lid for cracks or chips | Completed | |
| Inspect vacuum chamber seal for tears or cracks | Completed | |
| Confirm chamber lid micro switches report .04 Ohms | Completed | |
| Verify pump or cold trap connections are sealed and leak-free | Completed | |
| Inspect and clean internal fan and condenser (if applicable) | Completed | |
| Inspect internal wiring for frays or breakages | Completed | |
| Test motor functionality, replace motor/bearings if needed | Completed | |
| Inspect internal pump seals, replace if needed | Completed | |
| Inspect all internal hoses for leaks, ensure clean seals | Completed | |
| Confirm unit achieves specified temperature ranges | Completed | |
| Verify unit reaches specified vacuum levels | Completed | |
| Conduct a 6-8 hour cycle test for operational performance | Completed |