Eppendorf 5418 Centrifuge w/ FA-45-18-11 Rotor & Lid
Max RCF / RPM
16,873 x g / 14,000 RPM
Max Capacity
18 x 1.5/2.0ml
Electrical Requirements
120V, 50/60Hz, 170W
- Compact and easy-to-use, ideal for standard molecular biology and clinical applications.
- Capacity of 18 x 1.5/2.0 ml tubes and reaches a maximum of 16,873 x g / 14,000 rpm.
- Quiet operation with soft-touch one-finger lid closure, and ECO power shut-off feature for energy efficiency.
- Eppendorf QuickLock lid for ergonomic, secure locking with a quarter turn.
- Aerosol-tight containment for safe centrifugation of hazardous samples.
- Includes NIST Centrifuge Calibration and Preventative Maintenance Report.
The Eppendorf 5418 Centrifuge is specifically designed for low to medium capacity with standard molecular biology and clinical applications. The 5418 has a capacity of 18 x 1.5/2.0 ml tubes with speeds up to 14,000 RPM with a maximum RCF of 16,873 x g. This centrifuge offers a small footprint, low access height, quiet operation, soft-touch one-finger lid closure, and ECO power shut-off. Eppendorf QuickLock lid allows for ergonomic operation with fast and secure locking with only a quarter turn. Aerosol-tight containment allows safe centrifugation of hazardous samples.
The Eppendorf 5418 Centrifuge includes:
- > FA-45-18-11 Rotor & Lid
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Rotor Wrench
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5418 Centrifuge w/ FA-45-18-11 Rotor Specifications
|
Speed Range |
14,000 RPM |
|
Max RCF |
16,873 x g |
|
Max Capacity |
18 x 1.5/2.0ml |
|
Max Load |
3.75 g |
|
Acceleration Time |
16 s |
|
Deceleration Time |
18 s |
|
Timer |
30 s to 9:59 h, with continuous run function |
|
Dimensions (W x D x H) |
8.19 x 11.8 x 8.27 in |
|
Electrical Requirements |
120V, 50/60Hz, 170W |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Eppendorf 5418 Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly



Click Here to View the Technical Specs for the 5418