Eppendorf BioPhotometer Plus Model #6132
Measuring Wavelengths
230, 260, 280, 340, 405, 490, 550, 595, 650 nm
Wavelength selection
Method-dependent, program controlled
Wavelength Systematic Error
± 1 nm at 230 to 280 nm
- Single-beam photometer with reference beam for reliable and accurate measurements
- Fixed wavelengths for key research methods: 230, 260, 280, 340, 405, 490, 550, 595, and 650 nm
- High-precision spectral dispersion using a holographic concave grating
- Xenon flash lamp for consistent and durable irradiation
- Exceptional photometric accuracy with systematic error ±1% at 1 A
- 32 preprogrammed and modifiable method programs for versatile applications
- Compact and lightweight design with dimensions of 20 x 32 x 10 cm and weight of 3 kg
- Robust metal housing for stability and operational reliability
The Eppendorf BioPhotometer Plus Model #6132 includes:
- > Photometric Calibration Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf BioPhotometer Model #6132 Specifications
|
Optical system |
Absorption single-beam photometer with reference beam and several fixed wavelengths |
|
Irradiation source |
Xenon flash lamp |
|
Spectral dispersion |
Holographic concave grating |
|
Measuring wavelengths |
230, 260, 280, 340, 405, 490, 550, 595, 650 nm |
|
Wavelength selection |
Method-dependent, program controlled |
|
Spectral bandwidth |
5 nm at 230 to 340 nm 7 nm at 405 to 650 nm |
|
Wavelength systematic error |
± 1 nm at 230 to 280 nm ± 2 nm at 340 to 650 nm |
|
Photometric random error |
≤ 0.002 A at 0 A ≤ 0.005 A at 1 A |
|
Photometric systematic error |
± 1% at 1 A |
|
Accuracy of reading |
0.001 A |
|
Stray-light component |
< 0.05% |
|
Memory |
32 preprogrammed, modifiable method programs |
|
Supply voltage |
100 to 240 V ± 10%: 50 to 60 Hz |
|
Dimensions (W x D x H) |
20 x 32 x 10 cm |
|
Weight |
3 kg |
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm stable, correct voltage outputs from power supply and PCB | Completed | |
| Inspect display for dim or dead pixels; replace as needed | Completed | |
| Verify functionality of all buttons for precise, responsive control | Completed | |
| Check XY assembly for proper movement; clean railings to remove debris | Completed | |
| Confirm stable software communication between instrument and PC | Completed | |
| Inspect and clean or replace filters to maintain optimal performance | Completed | |
| Examine internal wiring for damage; replace or secure compromised connections | Completed | |
| Test lamp functionality; replace if dimming or faults are detected | Completed | |
| Execute validation protocol and document on calibration/verification certificate | Completed | |
| Update PC operating system for compatibility and security | Completed | |
| Conduct cosmetic inspection to ensure quality of external components | Completed |
Downloads


Click Here to View the Technical Specs for the BioPhotometer Plus