Eppendorf Research Pipette Set, 0.1 to 1000uL
# of Pipettes Included
Six
Volume Range
0.1 to 1000uL
Number of Channels
Single
- Lightweight construction minimizes fatigue during prolonged pipetting
- Autoclavable components to ensure ease of use in maintaining proper sterilization
- Low operating and tip ejection forces to reduce hand strain
- Three-button operation for simple volume setting, pipetting, and tip ejection
- Built according to the Eppendorf PhysioCare Concept for ergonomic comfort
- Durable, UV- and chemical-resistant materials for long-lasting performance
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf Research Pipette Set, 0.1 to 1000uL includes:
- > Six Eppendorf Research Pipettes: 0.1 to 2.5uL, 0.5 to 10uL, 2 to 20uL, 10 to 100uL, 20 to 200uL, and 100 to 1000uL
- > Gravimetric Analysis Calibration Reports
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Eppendorf Research Pipette Set, 0.1 to 1000uL Specifications
| Pipette Model | Volume (uL) | Systematic Error (uL) | Random Error (uL) |
| 0.1 to 2.5uL |
0.25 1.25 2.5 |
± 0.03 ± 0.031 ± 0.035 |
≤ 0.015 ≤ 0.019 ≤ 0.018 |
| 0.5 to 10uL |
1 5 10 |
± 0.025 ± 0.075 ± 0.1 |
≤ 0.018 ≤ 0.04 ≤ 0.04 |
| 2 to 20uL |
2 10 20 |
± 0.1 ± 0.12 ± 0.2 |
≤ 0.03 ≤ 0.06 ≤ 0.06 |
| 10 to 100uL |
10 50 100 |
± 0.3 ± 0.5 ± 0.8 |
≤ 0.1 ≤ 0.15 ≤ 0.2 |
| 20 to 200uL |
20 100 200 |
± 0.5 ± 1.0 ± 1.2 |
≤ 0.14 ≤ 0.3 ≤ 0.4 |
| 100 to 1000uL |
100 500 1000 |
± 3.0 ± 5.0 ± 6.0 |
≤ 0.6 ≤ 1.0 ≤ 2.0 |
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Thoroughly inspect and clean all internal components; replace damaged parts | Completed | |
| Replace essential wear components like O-rings, seals, and friction rings | Completed | |
| Examine plunger and seal holders for wear or damage; replace if needed | Completed | |
| Verify no leakage is present by aspirating sample and checking for fluid loss | Completed | |
| Confirm Mettler Toledo AB104-S balance calibration for accurate measurements | Completed | |
| Perform detailed gravimetric calibration and record findings in report | Completed | |
| Clean all external parts meticulously to remove any contaminants | Completed | |
| Test tip ejector for smooth, reliable operation; correct misalignments | Completed | |
| Affix dated calibration sticker to indicate verified calibration | Completed | |
| Conduct cosmetic inspection to ensure quality of surfaces and finishes | Completed |


Click Here to View the Technical Specs for the Eppendorf Research Pipette Set, 0.1 to 1000uL