Fisher Scientific Digital Dry Bath, catalog # 11-715-125D
Temperature Range
Ambient +5°C to 130°C
Temperature Uniformity
± 0.2°C
Block Capacity
2 Heat Blocks
- Accommodates Two Interchangeable Blocks for versatile sample handling
- Compatible with Tube Sizes from 0.25mL to 15mL for broad application use
- Thermometer Well in Each Block for precise temperature monitoring
- Close Tube-and-Block Contact enhances heat transfer efficiency
- Low-Density Heater Design ensures consistent performance
- Compact and Versatile Construction for easy integration into lab workflows
All of our equipment is fully tested before being shipped.
The Fisher Scientific Digital Dry Bath, catalog # 11-715-125D includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
Fisher Scientific Digital Dry Bath, catalog # 11-715-125D
| Temperature Range | Ambient +5°C to 130°C |
| Temperature Uniformity | ± 0.2°C |
| Block Capacity | 2 |
| Electrical Requirements | 120V, 50/60Hz, 1.25A |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |