GE AKTA UV-900 Detector
Wavelength Range
190–700 nm
Drift
< 2x10-4 AU/h at 254 nm
Maximum Flow Rate
100 ml/min
- Up to 3 wavelengths can be monitored simultaneously.
- Utilizes cutting-edge fiber optic technology to monitor up to three wavelengths simultaneously in the 190 to 700 nm range.
- Novel flow cell design and fiber-optic technology contribute to high sensitivity and an impressive signal-to-noise ratio, with minimal band broadening for accurate detection.
- Equipped with a Xenon flash lamp that offers high sensitivity, a continuous spectrum of light, for reliable and consistent performance.
- Contains a long path length with a small volume, designed to prevent the formation of lens-shaped interfaces between eluant components of different refractive indexes.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The GE AKTA UV-900 Detector includes:
- > New Lamp
- > Flow Cell
- > Installation Calibration Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
GE AKTA UV-900 Detector Specifications
|
Wavelength range
|
190–700 nm in steps of 1 nm,
3 wavelengths simultaneously
|
|
Bandwidth
|
4 nm
|
|
Wavelength accuracy
|
± 2 nm
|
|
Wavelength reproducibility
|
± 0.01 nm |
|
Linearity
|
< 2% deviation up to 2 AU at 260 nm with
Uracil at pH 2
|
|
Static noise
|
short term: < 6x10-5 AU at 230 nm
long term: < 6x10-5 AU at 230 nm
|
|
Drift
|
< 2x10-4 AU/h at 254 nm
|
|
Flow Cell
|
|
|
Optical path length 2 mm cell
|
2 mm, 2 µl cell volume
|
|
Optical path length 10 mm cell
|
10 mm, 8 µl cell volume
|
|
Max. flow rate
|
100 ml/min
|
|
Max. pressure
|
2 MPa (290 psi)
|
|
Temperature range
|
+4 to +40 °C (liquid)
|
|
Wetted material
|
PTFE (polytetrafluoroethylene)
PEEK (polyetheretherketone)
titanium (palladium alloy),
quartz (synthetic fused silica)
and titanium
|
|
Physical Data
|
|
|
Light source
|
Xenon flash lamp
|
|
Lamp lifetime
|
> 4000 hours
|
|
Control
|
Stand alone or from a computer running UNICORN 3.21 or higher through
UniNet-1 connection
|
|
Power consumption
|
65 VA
|
|
Power requirement
|
100–240 V AC, 50–60 Hz
|
|
Analogue output
|
3 signals, 0–1 V full scale, overrange function
|
|
Digital inputs
|
5 V, 1 mA current sinking, lamp on/off,
autozero, event mark
|
|
Display
|
2 rows with 20 characters each
|
|
Dimensions (H x W x D)
|
200 x 260 x 370 mm
|
|
Weight
|
8.5 kg
|
Downloads


Click Here to View all the Technical Specs for the GE AKTA UV-900 Detector