Gilson P1000 Pipette Classic Large Plunger
Adjustable Volume Range
100-1000 µL
Channels
1
Pipette Type
Manual Air Displacement
- The P1000 covers a volume range from 100 µL to 1000 µL, allowing flexibility for various sample types.
- With an ergonomic design, it ensures comfortable operation even during prolonged use, reducing user fatigue.
- Made of PVDF and stainless steel, the Pipetman P1000 offers premium durability for years of pipetting.
- Routine cleaning and maintenance keep the Pipetman in top condition. It has user-serviceable parts and a fully autoclavable tip holder and ejector.
- Built-to-last and known for accuracy and precision, the Pipetman P1000 is it is the economical solution for years of pipetting
Gilson PIPETMAN P1000 Pipette is a variable volume air displacement pipette for dispensing volumes up to 1000 ul for samples with precision accuracy and ease. Ideal for small-volume dispensing of aqueous fluids of moderate viscosity and density within immunology, molecular biology, etc. applications.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Gilson P1000 Pipette Classic Large Plunger includes:
- > Gravimetric Analysis Calibration
- > Full 180 Day Parts and Labor Warranty
Gilson P1000 Pipette Classic Large Plunger Specifications
|
Volume µl |
Increment µl |
Relative Accuracy (%) |
Absolute Accuracy (µl) |
Relative Precision (%) |
Absolute Precision (%) |
|
100 500 1000 |
2 2 2 |
3.0 0.8 0.8 |
3 4 8 |
0.6 0.2 0.15 |
0.6 1 1.5 |
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Thoroughly inspect and clean all internal components; replace damaged parts | Completed | |
| Replace essential wear components like O-rings, seals, and friction rings | Completed | |
| Examine plunger and seal holders for wear or damage; replace if needed | Completed | |
| Verify no leakage is present by aspirating sample and checking for fluid loss | Completed | |
| Confirm Mettler Toledo AB104-S balance calibration for accurate measurements | Completed | |
| Perform detailed gravimetric calibration and record findings in report | Completed | |
| Clean all external parts meticulously to remove any contaminants | Completed | |
| Test tip ejector for smooth, reliable operation; correct misalignments | Completed | |
| Affix dated calibration sticker to indicate verified calibration | Completed | |
| Conduct cosmetic inspection to ensure quality of surfaces and finishes | Completed |