Gilson PIPETMAN L Pipette, 0.5 to 10uL
Volume Range
0.5 to 10uL
Pipette Type
Manual Air Displacement
Number of Channels
Single
- Ergonomic design suitable for both right- and left-handed users
- Reduced pipetting forces minimize user fatigue
- Volume-locking mechanism prevents accidental adjustments
- Variable volume models are fully autoclavable
- 2D DataMatrix code for simplified traceability and asset management
- Unique serial number engraved for easy identification
- Highly durable construction ensuring long-term reliability
Traceability is built in, with a 2D DataMatrix code and unique serial number engraved on the body for GLP compliance and simplified asset management. Constructed for long-term performance, the PIPETMAN L Series is available in fixed and variable volume formats, including single and multichannel configurations to suit every lab’s needs.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Gilson PIPETMAN L Pipette, 0.5 to 10uL includes:
- > Gravimetric Analysis Calibration Report
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Gilson PIPETMAN L Pipette, 0.5 to 10uL Specifications
|
Volume uL |
Increment uL |
Relative Accuracy (%) |
Absolute Accuracy (uL) |
Relative Precision (%) |
Absolute Precision (uL) |
|
0.5 1 5 10 |
0.02 0.02 0.02 0.02 |
± 8.0 ± 2.5 ± 1.5 ± 1.0 |
± 0.040 ± 0.025 ± 0.075 ± 0.100 |
≤ 3.2 ≤ 1.2 ≤ 0.6 ≤ 0.4 |
≤ 0.016 ≤ 0.012 ≤ 0.030 ≤ 0.040 |
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Thoroughly inspect and clean all internal components; replace damaged parts | Completed | |
| Replace essential wear components like O-rings, seals, and friction rings | Completed | |
| Examine plunger and seal holders for wear or damage; replace if needed | Completed | |
| Verify no leakage is present by aspirating sample and checking for fluid loss | Completed | |
| Confirm Mettler Toledo AB104-S balance calibration for accurate measurements | Completed | |
| Perform detailed gravimetric calibration and record findings in report | Completed | |
| Clean all external parts meticulously to remove any contaminants | Completed | |
| Test tip ejector for smooth, reliable operation; correct misalignments | Completed | |
| Affix dated calibration sticker to indicate verified calibration | Completed | |
| Conduct cosmetic inspection to ensure quality of surfaces and finishes | Completed |


Click Here to View the Technical Specs for the Gilson PIPETMAN L Pipette, 0.5 to 10uL