Hermle Z216 M Centrifuge
Maximum Speed
15,000 RPM
Maximum Capacity
44 x 1.5/2.0ml
Max RCF
21,631 x g
- Operates as low as 200 RPM for delicate sample handling
- Versatile timer settings ranging from 10 seconds to 99 hours for precise control
- Compact design minimizes lab space usage while delivering maximum functionality
- Advanced microprocessor with user-friendly LCD interface for effortless operation
- Pre-cooling feature ensures samples reach optimal temperatures quickly and reliably
- Customizable acceleration and deceleration rates for enhanced sample protection
- Quiet operation with noise levels under 60 dBA for a more comfortable lab environment
Equipped with a user-friendly microprocessor and LCD interface, the Z216 M offers precise control over speed and time, with timer settings ranging from 10 seconds to 99 hours. The intuitive front panel allows effortless adjustment of acceleration and deceleration rates, with 10 customizable options for enhanced sample protection. Quiet operation, with noise levels below 60 dBA, ensures a comfortable lab environment.
The Hermle Z216 M features a durable, maintenance-free induction drive and advanced safety measures, including an active imbalance detection system and motor-driven lid lock. Engineered to meet the demands of modern molecular biology laboratories, this centrifuge is the perfect combination of power, versatility, and reliability.Please Note: A Rotor is NOT included with this centrifuge and is required for operation.
The Hermle Z216 M Centrifuge includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
- > Please Note: A Rotor is NOT included with this centrifuge and is required for operation
Hermle Z216 M Centrifuge Specifications
| Maximum Speed | 15,000 RPM |
| Minimum Speed | 200 RPM |
| Maximum RCF | 21,631 x g |
| Maximum Capacity | 44 x 1.5/2.0ml |
| Timer | 10 sec. to 99 hrs., 59 sec., or continous |
| Acceleration Rates | 10 |
| Deceleration Rates | 10 |
| Dimensions (W x H x D) | 11 x 11.4 x 15.3 in. (28 x 29 x 39 cm) |
| Weight | 37.4 lb. (17 kg) |
| Electrical Requirements | 120V, 50/60Hz, 2.3A |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Hermle Z216 MK Refrigerated Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly




Click Here to View the Technical Specs for the Hermle Z216 M Centrifuge