Hettich Universal 320 Centrifuge
- Capable of reaching up to 15,000 RPM and 21,382 x g for high-performance centrifugation
- Timer functionality supports runs up to 99 minutes with 10 programmable memory options
- Comprehensive safety features, including powered lid locking, lid dropping protection, and imbalance switch-off
- Compact dimensions of 346 x 395 x 520 mm for convenient benchtop use
- Operates on 100-127V, 50/60Hz for easy laboratory integration
- Non-refrigerated design suitable for a wide range of applications
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Hettich Universal 320 Centrifuge includes:
- > Hettich 1324 Swing Out Rotor and 1398 Buckets
- > 2 x Sets of 1484 50ml Adapters (4 Total)
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Hettich Universal 320 Centrifuge, Cat # 1401-01 Specifications
|
Maximum Speed |
15,000 RPM (fixed angle rotor) |
|
Maximum RCF |
21,382 x g |
|
Timer |
Up to 99 minutes |
|
Dimensions (H x W x D) |
346 x 395 x 520 mm |
|
Electrical Requirements |
100-127V, 50/60Hz |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Hettich Universal 320 Centrifuge, Cat # 1401-01
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly