Mettler Toledo XS204 Analytical Balance
Maximum Capacity
220 g
Readability
0.1 mg
Repeatability
0.1 mg
- SmartGrid weighing pan design minimizes air turbulence for faster stabilization and consistent results
- Glass draft shield can be completely dismantled for quick and thorough cleaning
- All components are dishwasher-safe for easy maintenance
- FACT automatic internal adjustment compensates for temperature fluctuations to ensure continuous accuracy
- Backlit touchscreen display provides intuitive operation and clear visibility of weighing parameters
- RS232 connectivity allows simple data transfer and system integration
- Compact, chemical-resistant design ideal for analytical, pharmaceutical, and QC applications
- Ergonomic layout enables easy access to the weighing chamber for efficient sample handling

Designed for simplicity, durability, and easy maintenance, the glass draft shield can be completely dismantled for convenient access, while all components are dishwasher-safe up to 80°C for effortless cleaning. The FACT automatic internal adjustment system continuously compensates for temperature variations, maintaining calibration accuracy without the need for manual intervention.
Equipped with a backlit touchscreen display for clear visibility and intuitive operation, the XS204 also includes RS232 connectivity for straightforward data transfer and integration with laboratory systems. Compact and robust, it offers the essential features needed for efficient daily use, combining accuracy, ease of operation, and low maintenance—making it a dependable choice for analytical, pharmaceutical, and quality control laboratories.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Mettler Toledo XS204 Analytical Balance includes:
- > Draft Shield
- > Calibration Label
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Mettler Toledo XS204 Analytical Balance Specifications
| Maximum Capacity | 220 g |
| Readability | 0.1 mg |
| Repeatability | 0.1 mg |
| Linearity | ± 0.2 mg |
| Calibration Type | Internal (Automatic / FACT) |
| Stabilization Time | 1.5 seconds |
| Minimum Weight (U=1%, k=2), Typical | 8.2 mg |
| Minimum Weight (USP, 0.1%, Typical) | 82 mg |
| Interfaces | RS232C, optional Ethernet, LocalCan, PS/2, Bluetooth and USB |
| Display | Backlit LCD touchscreen |
| Weighing Pan Size | 78 mm x 73 mm |
| Dimensions (H x W x D) | 10.4 x 17.8 x 12.7 in. (26.3 x 45.3 x 32.2 cm) |
| Weight | 20.1 lb. (9.1 kg) |
| Electrical Requirements | 100 to 240V, 50/60Hz |
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltage | Completed | |
| Assess display and control button functionality, replacing faulty components | Completed | |
| Confirm balance leveling for measurement accuracy | Completed | |
| Inspect windscreen glass for cracks, replace if necessary | Completed | |
| Check internal wiring and electrical connections for security | Completed | |
| Inspect power cord for frays or damage | Completed | |
| Execute internal/external calibration for measurement accuracy | Completed | |
| Review cosmetic condition for visible damage | Completed | |
| Verify weighing pan tares successfully | Completed | |
| Ensure RS232/printer communication is functional (if applicable) | Completed |


Click Here to View all the Technical Specs for the Mettler Toledo XS204 Analytical Balance