Maximum Capacity
220 g / 81 g
Readability
0.1 mg / 0.01 mg
Repeatability
0.1 mg / 0.04 mg
- FACT (Fully Automatic Calibration Technology) adjusts automatically based on a defined temperature criterion for consistent accuracy
- Internal calibration weight made from stainless, antimagnetic chrome-nickel steel for stable, interference-free adjustment
- Customizable FACT settings include temperature trigger and automatic report printing options
- SmartGrid weighing pan minimizes air turbulence for faster stabilization
- Ergonomic draft shield design provides easy access to the weighing chamber
- Touchscreen interface allows intuitive control and configuration
- Compact footprint conserves valuable bench space in the lab
- Durable construction ensures long-term reliability under routine laboratory use
The Mettler Toledo XS205DU incorporates stainless, antimagnetic chrome-nickel steel adjustment weights to ensure stable calibration over time and minimize magnetic interference. Its SmartGrid weighing pan design reduces air turbulence, providing faster stabilization and improved repeatability. The touchscreen interface offers intuitive operation, while the ergonomic draft shield provides easy access to the weighing chamber for effortless sample handling. Compact yet durable, the XS205DU’s construction and intelligent design make it ideal for analytical, pharmaceutical, and quality assurance applications requiring dependable precision and long-term stability.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Mettler Toledo XS205DU Semi-Micro Analytical Balance includes:
- > Calibration Label
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Mettler Toledo XS205DU Semi-Micro Analytical Balance Specifications
| Maximum Capacity | 220 g / 81 g |
| Readability | 0.1 mg / 0.01 mg |
| Repeatability | 0.1 mg / 0.04 mg |
| Linearity | 0.2 mg |
| Calibration Type | Internal (FACT) |
| Stabilization Time | 1.5 seconds |
| Minimum weight (according to USP) | 120 mg + 0.0006 x Rgr |
| Minimum weight (according to USP), fine range | 30 mg + 0.00012 x Rgr |
| Minimum weight (U=1%, 2 sd) | 8 mg + 0.00004 x Rgr |
| Minimum weight (U=1%, 2 sd) fine range | 2 mg + 0.00008 x Rgr |
| Display | Backlit touchscreen |
| Weighing Pan Dimensions (W x D) | 3 x 2.8 in. (7.8 x 7.3 cm) |
| Dimensions (W x D x H) | 10.4 x 17.8 x 12.7 in. (26.3 x 45.3 x 32.2 cm) |
| Usable Height of Draftshield | 9.3 in. (23.5 cm) |
| Weight | 20.1 lb. (9.1 kg) |
| Electrical Requirements | 100 to 240V, 50/60Hz, 0.8A |
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltage | Completed | |
| Assess display and control button functionality, replacing faulty components | Completed | |
| Confirm balance leveling for measurement accuracy | Completed | |
| Inspect windscreen glass for cracks, replace if necessary | Completed | |
| Check internal wiring and electrical connections for security | Completed | |
| Inspect power cord for frays or damage | Completed | |
| Execute internal/external calibration for measurement accuracy | Completed | |
| Review cosmetic condition for visible damage | Completed | |
| Verify weighing pan tares successfully | Completed | |
| Ensure RS232/printer communication is functional (if applicable) | Completed |