New Brunswick I26 Incubator Shaker
Speed Range
25 to 400 RPM (25 to 250 RPM for Stacked Units)
Temperature Range
Ambient +5C to 60C
Orbit Size
2.5 cm (1 in)
- Microprocessor controls precisely regulate temperature, speed and running time
- Slide out platform provides easy access to all samples, including ones located in the back
- User-friendly keypad and large LED display allows for clear visibility and easy set of speed and temperature
- Stack up to 3 units for maximum capacity while taking up minimal space
- Triple-eccentric counterbalanced drive and cast iron housing provides vibration and trouble free operation for years
The New Brunswick I26 Incubator Shaker is a great value option for a stackable orbital incubator shaker that combines a triple eccentric counterbalanced drive mechanism with precision temperature control, providing trouble-free operation for years. This cast iron triple-eccentric counterbalanced drive is specifically designed to support high-speed applications and heavy workloads. The I26 serves as the basic model of the Innova 44 Incubator Shaker series and features an easy-to-use folding door instead of the space-saving glide-up doors. A speed range of 25 to 400 RPM (25 to 250 RPM if stacking 2 or 3 units) is utilized with a temperature range of Ambient +5°C to 80°C. This M1324-0000 model of the I26 Incubator Shaker also features an orbit size of 2.5 cm (1 in). Microprocessor controls and a large digital display allow for easy operation and accurate setpoint regulation, along with the ability to set a constant speed and temperature or timed shaking.
The New Brunswick I26 Incubator Shaker includes:
- > 30" x 18" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick I26 Incubator Shaker Specifications
|
Speed Range
|
25 to 400 RPM (25 to 250 RPM for Stack of 2 to 3 Units)
|
| Speed Control Accuracy | ± 1 RPM |
|
Temperature Range
|
Ambient +5°C to 60°C |
|
Temperature Uniformity
|
± 0.25°C at 37C
|
| Timer | 0.1 to 99.9 hours, or continuous |
| Program Modes | Constant speed and temperature and timed shaking |
| Orbit Size | 2.5 cm (1 in) |
|
Platform Size
|
30 x 18 in (76 x 46 cm)
|
|
Chamber Dimensions (W x D x H)
|
34 x 23 x 15.5 in (86.4 x 58.4 x 39.4 cm) |
| Overall Dimensions (W x D x H) | 50 x 30.26 x 27.56 in (127 x 77 x 70 cm) |
| Weight | 370 lb. (167.8 kg) |
|
Electrical Requirements
|
120V, 60Hz, 1Ph, 1500VA |
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |


Click Here to View the Technical Specs for the New Brunswick I26 Incubator Shaker