New Brunswick I26 Incubator Shaker Dual Stack
Speed Range
25 to 250 RPM
Temperature Range
Ambient +5°C to 60°C
Orbit Size
2.5 cm (1 in)
- Microprocessor controls precisely regulate temperature, speed and running time
- Slide out platform provides easy access to all samples, including ones located in the back
- User-friendly keypad and large LED display allows for clear visibility and easy set of speed and temperature
- Stack up to 3 units for maximum capacity while taking up minimal space (currently includes a stack of 2 units)
- Triple-eccentric counterbalanced drive and cast iron housing provides vibration and trouble free operation for years
The New Brunswick I26 Incubator Shaker Dual Stack is a great value option for a dual stacked orbital incubator shaker that combines a triple eccentric counterbalanced drive mechanism with precision temperature control, providing trouble-free operation for years. This cast iron triple-eccentric counterbalanced drive is specifically designed to support high-speed applications and heavy workloads. The I26 serves as the basic model of the Innova 44 Incubator Shaker series and features an easy-to-use folding door instead of the space-saving glide-up doors. A speed range of 25 to 250 RPM is utilized with a temperature range of Ambient +5°C to 60°C. This M1324-0000 model of the I26 Incubator Shaker also features an orbit size of 2.5 cm (1 in). Microprocessor controls and a large digital display allow for easy operation and accurate setpoint regulation, along with the ability to set a constant speed and temperature or timed shaking.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The New Brunswick I26 Incubator Shaker Dual Stack includes:
- > 30" x 18" Platform Per Unit
- > Stacking Kit
- > NIST Calibration and Preventative Maintenance Report
- > Power Cords
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick I26 Incubator Shaker Dual Stack Specifications
|
Speed Range |
25 to 250 RPM |
| Speed Control Accuracy | ± 1 RPM |
|
Temperature Range |
Ambient +5°C to 60°C |
|
Temperature Uniformity |
± 0.25°C at 37C |
| Timer | 0.1 to 99.9 hours, or continuous |
| Program Modes | Constant speed and temperature and timed shaking |
| Orbit Size | 2.5 cm (1 in) |
| Platform Size |
30 x 18 in (76 x 46 cm) [per unit] |
|
Chamber Dimensions (W x D x H) |
34 x 23 x 15.5 in (86.4 x 58.4 x 39.4 cm) [per unit] |
| Overall Dimensions (W x D x H) | 50 x 30.26 x 27.56 in (127 x 77 x 70 cm) [per unit] |
| Weight | 740 lb. (335.6 kg) |
|
Electrical Requirements |
120V, 60Hz, 1Ph, 1500VA [per unit] |
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |