New Brunswick Innova 4300 Incubator Shaker
Temperature Range
Ambient +5C - 60C
Orbit
1 inch (25.4 mm)
Shaking Speed
25 - 500 RPM
- Spacious internal chamber measuring 34¾ inches wide, 22¼ inches deep, 19¼ inches high above the platform
- Accommodates flasks up to 6 liters
- Features a foot-pedal lid lift for convenience
- Operational modes: Continuous or timed with a programmable timer from 0.1 to 99.9 hours
- Audible and visible alarms: End of a timed run, Deviations in shaking speed or temperature beyond set limits
- Supports a variety of dedicated and universal platforms to accommodate different flask sizes and test tube racks
- Analog outputs available for connecting a chart recorder to track speed and temperature data
The New Brunswick Innova 4300 Incubator Shaker is a
large-capacity environmental shaker that allows for uniform heating and motion
and accurate regulation of shaking speed. This equilibration creates faster
operation and enhanced productivity for laboratories with shaking requirements. Designed
for handling greater quantities of flasks, the Innova 4300 easily holds 12 x 2
L flasks and accepts flasks up to 6 L. The 4300 shaker provides temperatures
from 5C above ambient to 60C for incubation. Shaking speeds from 25 RPM to 500 RPM on a 1 inch circular orbit. The 4300 features alarms for deviations of shaking speed, temperature outside of tolerance limits, and end of timed run.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The New Brunswick Innova 4300 Incubator Shaker includes:
- > 18 x 30" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 4300 Incubator Shaker Specifications
|
Platform Size
|
18 x 30 in.
|
|
Maximum Capacity (250 ml flasks on universal platform)
|
30
|
|
Capacity & Largest Flask Size on Universal Platform
|
4 x 6 L
|
|
Orbit Diameter
|
1 in
|
|
Temperature Range
|
Ambient +5C to 60C
|
|
Agitation (rpm)
|
25 to 500
|
|
Overall Dimensions (W x D x H)
|
43 x 28 x 37 in
|
|
Electrical Requirements
|
110-230V, 60Hz |
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |