Ohaus NV1101 Navigator Top-Loading Portable Balance
The Ohaus NV1101 Navigator Top-Loading Portable Balance sets a new standard for value and efficiency in weighing technologies. This balance boasts a maximum capacity of 1,100 g with a readability of 0.1 g, ensuring high precision for a wide range of applications. It features rapid stabilization in less than one second, enhancing productivity and throughput. Uniquely equipped with touchless sensors, the NV1101 allows for hands-free operation, minimizing contamination and facilitating ease of use during sample handling. These sensors can control up to 36 combinations of operations, including zero, print, function, and tare, making the balance ideal for busy laboratory environments.
Additionally, the NV1101 offers robust overload protection, capable of handling loads up to four times its rated capacity, which prevents damage from excessive loading and ensures long-term reliability. Its durable design includes a polished stainless steel platform and corrosion-resistant plastic housing, making it easy to clean and maintain. The balance also supports multiple application modes such as counting, percent weighing, and checkweighing, suitable for various tasks from quality control to inventory management. An internal rechargeable battery provides up to 100 hours of operation, enhancing its portability and use in areas without immediate access to power. The bright backlit LCD ensures visibility under any lighting conditions, further asserting the NV1101 as an indispensable tool for fast-paced settings requiring both precision and durability.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Ohaus NV1101 Navigator Top-Loading Portable Balance includes:
- > Calibration Label
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Ohaus NV1101 Navigator Top-Loading Portable Balance Specifications
| Maximum Capacity | 1,100 g |
| Readability | 0.0005 lb. (0.1 g) |
| Repeatability | 0.1 g |
| Linearity | ± 0.2 g |
| Calibration Type | User-selectable external span or linearity calibration / Digital with external weight |
| Stabilization Time | ≤ 1 second |
| Display | LCD with backlight |
| Application Modes | Weighing, Percent Weighing, Parts Counting, Check Weighing, Display Hold, Accumulate/Totalize |
| Weighing Pan Dimensions (W x D) | 7.5 x 5.4 in. (19 x 13.8 cm) |
| Overall Dimensions (W x D x H) | 8 x 8.4 x 2.3 in. (20.4 x 21.2 x 5.8 cm) |
| Weight | 3.1 lb. (1.4 kg) |
| Electrical Requirements | 12VDC 0.42A AC adapter |
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltage | Completed | |
| Assess display and control button functionality, replacing faulty components | Completed | |
| Confirm balance leveling for measurement accuracy | Completed | |
| Inspect windscreen glass for cracks, replace if necessary | Completed | |
| Check internal wiring and electrical connections for security | Completed | |
| Inspect power cord for frays or damage | Completed | |
| Execute internal/external calibration for measurement accuracy | Completed | |
| Review cosmetic condition for visible damage | Completed | |
| Verify weighing pan tares successfully | Completed | |
| Ensure RS232/printer communication is functional (if applicable) | Completed |