Temperature Range
Ambient +5°C to 50°C
CO2 Range
0 to 20%
Volume
170L
- Advanced humidity and temperature control technology for precise culture conditions
- Direct heat and air jacket system eliminates the need for water and provides even heat distribution
- Dual detector infrared CO2 sensor maintains constant conditions regardless of humidity or temperature changes
- Safe Cell UV contamination control system ensures a sterile environment
- Spacious interior with 170L volume and two stainless steel trays
- Rigid polyurethane foamed-in-place insulation for efficient temperature control
- PID control system for temperature and CO2 levels with digital displays
- Air circulation with fan assistance for uniform conditions
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack includes:
- > 2 x Shelves Per Unit
- > 2 x Humidity Pans
- > Stacking Kit
- > Calibration and Preventative Maintenance Report
- > 2 x Power Cords
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack Specifications
| Temperature Range | Ambient +5°C to 50°C |
| Temperature Uniformity | ± 0.25°C at 37°C |
| Temperature Stability | ± 0.1°C at 37°C |
| CO2 Range | 1 to 20% |
| CO2 Control Deviation | ± 0.15% |
| CO2 Sensor Type | IR |
| Required Gas Pressure | 0.03 MPaG (0.3 kgf/cm2 G, 4.3 psiG) |
| Relative Humidity | 95% ± 5% |
| Volume | 6 cu ft. (170L) [Per Unit] |
| Air Filter | 0.3 um, efficiency: 99.97% or higher |
| UV Lamp | 4W X 1 (ozone-free emission) |
| Interior Material | Stainless steel containing copper |
| Access Port Diameter | 30 mm |
| Alarms | Automatic set temperature alarm, automatic set CO2 density alarm, upper limit temperature alarm, door alarm, low humidifying water alarm, UV lamp failure alarm |
| Maximum Heat Emission | 1,120 kJ/h |
| Shelf Dimensions (W x D x H) | 17 x 17 x 4.7 in. (45 x 45 x 12 cm) |
| Internal Dimensions (W x D x H) | 19 x 20.6 x 26.2 in. (49 x 52.3 x 66.5 cm) [Per Unit] |
| External Dimensions (W x D x H) | 24.4 x 27.9 x 35.4 in. (62 x 71 x 90 cm) [Per Unit] |
| Weight | 205 lb. (93 kg.) [Per Unit] |
| Electrical Requirements | 110-120V, 60 Hz, 2.8A, 350W [Per Unit] |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, COâ‚‚, and Hâ‚‚O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate COâ‚‚ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and COâ‚‚ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, COâ‚‚, and Hâ‚‚O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate COâ‚‚ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and COâ‚‚ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |


Click Here to View the Technical Specs for the Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack