Sanyo MCO-20AIC CO2 Incubator
Temperature Range
Ambient +5°C to 50°C
CO2 Range
0 to 20 %
Internal Dimensions
24.4 x 20.6 x 26.2 in.
- Equipped with a UV contamination control system for maintaining a sterile environment.
- Infrared sensor ensures precise CO2 level control, vital for sensitive cultures.
- Microprocessor PID Control provides reliable and consistent environmental conditions.
- Outer door with a magnetic seal and a door heater to eliminate moisture.
- Top and rear ducts are removable for easy maintenance.
- Water Level Sensor monitors water levels in the humidifying pan, ensuring optimal humidity.
- Disinfectable Fan made from polypropylene resin and suitable for autoclave disinfection.
The Sanyo MCO-20AIC CO2 Incubator is designed to deliver precise humidity, CO₂, and temperature control for cell cultures in medical, pharmaceutical, and research laboratories while maintaining a sterile environment through its integrated UV contamination control system. An infrared sensor ensures accurate CO₂ level adjustments essential for sensitive cultures, and a microprocessor PID control provides reliable, consistent conditions. The outer door features a magnetic seal with a door heater to eliminate moisture, and both the top and rear ducts are removable for effortless maintenance. Additionally, a water level sensor monitors the humidifying pan to maintain optimal humidity, and a disinfectable fan made from polypropylene resin is suitable for autoclave disinfection, making this incubator a robust and user-friendly solution for advanced culturing needs.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Sanyo MCO-20AIC CO2 Incubator includes:
- > UV Lamp
- > 3 Shelves
- > Drain Tray
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Sanyo MCO-20AIC CO2 Incubator Specifications
| Temperature Range | Ambient +5°C to 50°C |
| Temperature Distribution | ± 0.25°C at 37°C (ambient temperature: 25C, setting: 37C, CO2: 5%, no load) |
| Temperature Variation | ± 0.1°C (ambient temperature: 25C, setting: 37C, CO2: 5%, no load) |
| Volume | 7.59 cu. ft. (215L) |
| CO2 Range | 0 to 20% |
| CO2 Variation | ± 0.15 % (ambient temperature: 25C, setting: 37C, CO2: 5%, no load) |
| Internal Dimensions (W x D x H) | 24.4 x 20.6 x 26.2 in. (62 x 52.3 x 66.5 cm) |
| Electrical Requirements | 110-120V, 60Hz, 3.7A, 380W |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, COâ‚‚, and Hâ‚‚O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate COâ‚‚ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and COâ‚‚ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, COâ‚‚, and Hâ‚‚O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate COâ‚‚ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and COâ‚‚ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |


Click Here to View the Technical Specs for the MCO-20AIC