Sartorius Cubis II MCA125P-2S00-U Semi-Micro Analytical Balance
Maximum Capacity
60 g/120 g
Readability
0.01 mg/0.1 mg
Repeatability
± 0.06 mg
- Designed for regulated R&D and analytical labs requiring pharmaceutical and GxP compliance
- Internal isoCAL automatically maintains calibration stability for long-term reliable performance
- Optimum USP Chapter 41 minimum weight of 8.2 mg supports compliant, low-mass weighing
- Large 7 in color TFT touchscreen with touch, scroll, and swipe for intuitive, modern operation
- Built-in applications and intelligent diagnostics guide users through workflows and reduce errors
- Fast stabilization time of 2 seconds for high-throughput labs needing rapid, repeatable results
- Extensive connectivity including USB-A, USB-B, USB-C, RS232, and Ethernet (LAN) for seamless system integration
- Ergonomic 85 mm x 85 mm weighing pan supports comfortable sample handling and method consistency
- End-to-end data integrity features support secure data handling and traceable documentation in regulated environments
- Seamless integration into digital lab ecosystems without additional middleware thanks to versatile interface options
The Sartorius Cubis II MCA125P-2S00-U Semi-Micro Analytical Balance is designed for scientists in R&D and analytical laboratories who demand uncompromising accuracy, reliability, and compliance in their daily weighing workflows. Built on the Cubis II platform, this balance offers a fully configurable combination of hardware, software, and connectivity so it can be tailored to your specific methods, SOPs, and regulatory requirements. Its customizable modularity allows you to adapt applications, accessories, and software over time as your laboratory needs evolve, while built-in weighing applications guide users step by step through even complex processes. This reduces training time, minimizes operator-to-operator variability, and supports consistent, traceable results across your entire lab.
At the heart of the Cubis II MCA125P-2S00-U is a powerful weighing system engineered for trusted, high-performance results. The balance offers a maximum capacity of 60 g/120 g with dual-range readability of 0.01 mg/0.1 mg, repeatability of ± 0.06 mg, and linearity of ± 0.15 mg, providing the precision needed for demanding semi-micro and analytical applications. Internal isoCAL automatic calibration maintains long-term stability and accuracy, while the optimum USP Chapter 41 minimum weight of 8.2 mg supports compliant weighing in regulated environments. Intelligent diagnostic systems, combined with optimized stabilization time of 2 seconds, help deliver high-speed, error-resistant measurements and a higher degree of repeatability across different workflows.
The Cubis II platform is also built to integrate seamlessly into modern digital and regulated laboratories. A large 7 in. color TFT touchscreen with high-resolution graphics, touch, scroll, and swipe functionality provides a modern user experience, making complex tasks intuitive and reducing the likelihood of input errors. The system supports full end-to-end data integrity with advanced data handling and connectivity options, enabling secure transfer of weighing data to external systems. Multiple interfaces, including USB-A, USB-B, USB-C, RS232, and Ethernet (LAN) allow direct, software-free integration with a wide range of laboratory information systems and external devices. Together with ergonomic sample handling, an 85 mm x 85 mm weighing pan, and easy process integration, the Cubis II MCA125P-2S00-U offers a future-ready weighing solution that aligns with stringent pharmaceutical and GxP requirements while maximizing operational efficiency and experimental confidence.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Sartorius Cubis II MCA125P-2S00-U Semi-Micro Analytical Balance includes:
- > Electronics Module w/ Connection Cable
- > Calibration Label
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Sartorius Cubis II MCA125P-2S00-U Semi-Micro Analytical Balance Specifications
| Maximum Capacity | 60 g/120 g |
| Readability | 0.01 mg/ 0.1 mg |
| Repeatability | ± 0.06 mg |
| Linearity | ± 0.15 mg |
| Calibration Type | Internal (isoCAL) |
| Stabilization Time | 2 seconds |
| Minimum Weight (USP, Chap. 41, Optimum) | 8.2 mg |
| Interfaces | USB-A, USB-B, USB-C, RS232, and Ethernet (LAN) |
| Display | 7 in. color TFT touchscreen |
| Weighing Pan Size | 85 mm x 85 mm |
| Electronics Module Dimensions (L x W x H) | 8.3 x 9.4 x 2.2 in. (21.1 x 24 x 5.6 cm) |
| Balance Dimensions (L x W x H) | 17.7 x 9.4 x 14.8 in. (45 x 24 x 37.5 cm) |
| Weight | 24.3 lb. (11 kg) |
| Electrical Requirements | 100 to 240V, 50/60Hz, 1.0A |
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltage | Completed | |
| Assess display and control button functionality, replacing faulty components | Completed | |
| Confirm balance leveling for measurement accuracy | Completed | |
| Inspect windscreen glass for cracks, replace if necessary | Completed | |
| Check internal wiring and electrical connections for security | Completed | |
| Inspect power cord for frays or damage | Completed | |
| Execute internal/external calibration for measurement accuracy | Completed | |
| Review cosmetic condition for visible damage | Completed | |
| Verify weighing pan tares successfully | Completed | |
| Ensure RS232/printer communication is functional (if applicable) | Completed |


Click Here to View all the Technical Specs for the Sartorius Cubis II MCA125P-2S00-U Semi-Micro Analytical Balance