Set of 6 x Cytoclips for Cytospin Centrifuges
This Set of 6 x Cytoclips are for use with the Shandon/Thermo Scientific Cytospin centrifuges and are designed to hold cytology funnels. These cytoclips are comprised of a stainless steel material.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Set of 6 x Cytoclips for Cytospin Centrifuges includes:
- > 6 x Cytoclips
- > Full 180 Day Parts and Labor Warranty
Set of 6 x Cytoclips for Cytospin Centrifuges Specifications
| Material |
Stainless Steel |
| For Use With | Shandon/Thermo Scientific Cytospin Centrifuges |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |