Sorvall Legend RT Refrigerated Benchtop Centrifuge
Maximum Speed
15,000 RPM
Maximum Capacity
4 x 750 mL
Temperature Range
-9°C to 40°C
- User-friendly with an easy-to-read digital display and simple control of centrifuge capabilities
- Low height design of 12.4 inches for convenient rotor loading and unloading
- Offers a substantial maximum g-force of approximately 24,400 x g
- Features a versatile timer setting ranging from 1 minute to 9 hours and 59 minutes
- Refrigerated benchtop design, making it ideal for temperature-sensitive applications
- Suitable for a broad range of laboratory situations, enhancing its versatility and utility
The Sorvall Legend RT Refrigerated Benchtop Centrifuge
offers remarkable user-friendly capabilities, including an easy-to-read
digital display and easy user control of all centrifuge capabilities. With a
low height of only 12.4 inches, the Sorvall Legend RT makes rotor loading and
unloading effortless. This refrigerated benchtop centrifuge has a max g-force of about 24,400 x
gs and features a timer that can be set at points as low as 1 min all the way to time periods lasting up to 9 hours and 59 minutes. With a temperature
range of -9 to +40C, the Sorvall Legend RT is the perfect centrifuge for any
laboratory situation.
Max speed with this rotor is 4,150 RPM.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Sorvall Legend RT Refrigerated Benchtop Centrifuge includes:
- > 75006434 Swing Bucket Rotor and Buckets
- > 1 Set of 15ml Adapters
- > 1 Set of 50ml Adapters
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Sorvall Legend RT Refrigerated Benchtop Centrifuge Specifications
| Maximum Speed | 15,000 RPM (4,150 RPM with included rotor) |
| Maximum RCF | 24,400 x g (3,696 x g with included rotor) |
| Maximum Capacity | 4 x 750ml |
| Temperature Range | -9C to +40C |
| Timer | 1 minute to 9 hours, 59 minutes |
| Dimensions (W x D x H) | 28.7 x 26.2 x 14.3 in. (11.3 x 66.5 x 36.3 cm) |
| Weight | 288 lb. (130.6 kg) |
| Electrical Requirements | 120V, 50/60Hz |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Sorvall Legend RT Refrigerated Benchtop Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly



Click Here to View the Technical Specs for the Legend RT