Thermo Fisher Scientific Ceramic Cimarec+ Hot Plate Stirrer
The Thermo Fisher Scientific Ceramic Cimarec+ Hot Plate Stirrer (Cat # SP88857100) offers users precise stirring control as well as incomparable safety performance. Advanced technology allows for slow speed stirring, consistent speed control, and enhanced magnetic coupling. The Thermo Fisher Scientific Ceramic Cimarec+ Top Stirrer has a speed range of 50 to 1,500 RPM.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Fisher Scientific Ceramic Cimarec+ Hot Plate Stirrer includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Fisher Scientific Ceramic Cimarec+ Hot Plate Stirrer Specifications
|
Temperature Range |
Ambient to 540C |
|
Stirrer Speed |
50 to 1,500 RPM |
|
Platform Size |
7.25 x 7.25 in (18.4 x 18.4 cm) |
|
Electrical Requirements |
120V, 50/60Hz |
| Top Plate Material | Ceramic |
| Overall Dimensions | 13 x 8.2 x 3.8 in (33 x 20.8 x 9.7 cm) |
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, COâ‚‚, and Hâ‚‚O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate COâ‚‚ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and COâ‚‚ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, COâ‚‚, and Hâ‚‚O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate COâ‚‚ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and COâ‚‚ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed |
Click Here to View all the Technical Specs for the Thermo Cimarec+ Top Stirrer


Click Here to View the Technical Specs for the Cimarec+ Top Hotplate Stirrer