Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer
Maximum Temperature
400C
Stirrer Speed
50 to 1,500 RPM
Platform Size
10.25 x 10.25 in (26.7 x 26.7 cm)
- Temperature adjustable in precise 1-degree increments for accurate control
- HOT TOP warning system alerts users when the surface exceeds 50C to prevent accidental burns
- Automatic over-temperature protection enhances safety during operation
- Durable ceramic top plate offers uniform heat distribution
- Easy-to-read LED display provides clear visibility of heating and stirring settings
- Raised display design safeguards electronics from spills and splashes
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer Specifications
| Maximum Temperature | 400C |
| Stirrer Speed | 50 to 1,500 RPM |
| Max. Stirring Volume | 6L |
| Platform Size | 10.25 x 10.25 in (26.7 x 26.7 cm) |
| Plate Material | Ceramic |
| Weight | 13.0 lb. (5.9 kg) |
| Dimensions (W x D x H) | 11.3 x 17.2 x 4.1 in. (28.8 x 43.8 x 10.5 cm) |
| Electrical Requirements | 100-120V, 50/60Hz, 13.2A |
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, COâ‚‚, and Hâ‚‚O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate COâ‚‚ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and COâ‚‚ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, COâ‚‚, and Hâ‚‚O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate COâ‚‚ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and COâ‚‚ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed |


Click Here to View the Technical Specs for the Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer