Speed Range
150 to 1,200 RPM
Speed Accuracy
± 15 RPM
Microplate Capacity
4 x 96-well microplates
- Compact design allows easy integration into incubators, chambers, and refrigerators
- PID control ensures precise and consistent shaking for reproducible experimental results
- Vacuum fluorescent display (VFD) provides clear and bright visibility of operational settings
- Adjustable timer allows for precise timing from 1 minute to 99 hours and 59 minutes
- Robust construction supports a maximum load of 8.8 lbs (4 kg) for diverse applications
- 3 mm orbit provides uniform motion, ideal for gentle yet effective mixing
- Automatic overload shutdown feature enhances safety and prolongs equipment lifespan
- Sturdy, space-saving design optimizes benchtop efficiency without compromising performance
- User-friendly interface allows for quick setup and intuitive operation
The shaker offers a broad speed range from 150 to 1,200 RPM with an accuracy of ±15 RPM, and its timer can be set from 1 minute to 99 hours and 59 minutes or operated continuously. It features an orbit of 3 mm and is built to handle a maximum load of 8.8 lbs (4 kg), accommodating up to four 96-well microplates. The platform measures 10.5 x 7.3 inches (26.7 x 18.5 cm), and the unit includes a VFD (vacuum fluorescent display) for clear visibility. Additionally, it is equipped with an automatic overload shutdown feature for added safety and durability.
The Thermo Scientific Compact Digital Microplate Shaker, Cat. #: 88880023 includes:
- > 10.5" x 7.3" Microplate Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Compact Digital Microplate Shaker, Cat. #: 88880023 Specifications
| Speed Range | 150 to 1,200 RPM |
| Speed Accuracy | ± 15 RPM |
| Timer | 1 minutes to 99 hours 59 minutes or continuous |
| Orbit |
3 mm (0.12 in.) |
| Display |
VFD (vacuum fluorescent display) |
| Maximum Load |
8.8 lb. (4 kg) |
| Microplate Capacity |
4 x 96-well microplates |
| Platform Size (W x D) |
10.5 x 7.3 in. (26.7 x 18.5 cm) |
| Dimensions (W x D x H) |
9.8 x 13 x 7.3 in. (24.9 x 32.9 x 12.7 cm) |
| Weight |
17 lb. (7.7 kg) |
| Electrical Requirements | 100-240V, 50/60Hz |
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Validate stable, correct voltage outputs from power supply and PCB | Completed | |
| Inspect display for dim, flickering, or dead pixels; replace as needed | Completed | |
| Test button functionality for responsiveness and accuracy across all settings | Completed | |
| Inspect bearings and housing for wear; replace as needed to prevent disruption | Completed | |
| Examine drive belt for cracking, fraying, or material breakdown; replace as required | Completed | |
| Measure and adjust motor drive belt tension to optimal specifications | Completed | |
| Verify onboard timer accuracy; replace if it deviates from expected performance | Completed | |
| Calibrate shaker rotation speed using a NIST-traceable tachometer | Completed | |
| Run 8-12 hour continuous test; observe platform for smooth, consistent motion | Completed | |
| Perform comprehensive cosmetic inspection for quality standards compliance | Completed |


Click Here to View the Technical Specs for the Thermo Scientific Compact Digital Microplate Shaker, Cat. #: 88880023