Thermo Scientific Heracell VIOS 160i CO2 Incubator
CO2 Range
1 to 20%
Capacity
5.8 cu ft (165 L)
Temperature Range
3°C above ambient to 55°C
- Designed to deliver the performance reliability, ease of operation, and value to support a range of culturing needs from basic to advanced applications
- THRIVE Active Airflow ensures even air distribution and fast recovery of all parameters ten minutes or less
- Auto-stop function disables fan operation to minimize air exchange during door openings
- ISO 5 HEPA filtration continuously filters air every 60 seconds and can achieve ISO 5 clean room air quality in 5 minutes
- Protected humidity reservoir prevents culture desiccation and contamination with 5X faster humidity recovery than traditional water pan designs
- Dual temperature probes for over-temperature protection
- Steri-Run cycle heats to 180°C for uniform 6-log sterilization on all chamber surfaces
- iCAN™ touchscreen control allows for easy accessibility and viewing
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific Heracell VIOS 160i CO2 Incubator includes:
- > 2 Shelves Per Unit
- > Power Cord
- > Calibration Report
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Heracell VIOS 160i CO2 Incubator Specifications
| Temperature Range | Ambient +3°C to 55°C |
| Temperature Uniformity | ± 0.1°C @ +37°C |
| Chamber Material | Electropolished stainless steel |
| Disinfection Time | Under 12 hours |
| CO2 Concentration Range | 1 to 20% |
| CO2 Sensor Technology | TC sensor |
| CO2 Concentration Control | ± 0.1% |
| Humidity Delivery | Integrated reservoir |
| Relative Humidity | >93% @ 37°C |
| Capacity | 5.8 cu ft (165 L) |
| Weight | 183 lb. (83 kg) |
| Interior Dimensions (H x D x W) | 23.9 x 22.7 x 18.5 in (607 x 576 x 470 mm) |
| Exterior Dimensions (H x W x L) | 35.6 x 25.1 x 31.1 in (905 x 637 x 790 mm) |
| Electrical Requirements | 120V, 50/60Hz, 4.6A/8.3A |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, COâ‚‚, and Hâ‚‚O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate COâ‚‚ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and COâ‚‚ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, COâ‚‚, and Hâ‚‚O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate COâ‚‚ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and COâ‚‚ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |
Downloads


Click Here to View the Technical Specs for the Heracell VIOS 160i