Temperature Range
-30°C Preset
Volume
29.2 cu. ft. (826L)
Defrost Method
Automatic
- Forced air circulation ensures uniform conditions and rapid recovery
- Adjustable epoxy-coated wire shelves with optional drawer configuration
- Self-closing door with a 90° stop for secure storage
- Keyed on/off switch with setpoint security to prevent unauthorized changes
- Remote alarm contacts and optional 4-20 milliamp output for monitoring
- Icon-based control panel with a digital display for precise adjustments
- Audible and visual alarms with a push-button test function
- Heavy-duty casters with front locks for easy mobility and stability
User convenience is addressed through its icon-based control panel and a large, easy-to-read digital display that shows temperature variations within 1°C. The graphic thermometer indicator provides visual alerts for overtemperature and under-temperature conditions. Additional features include audible/visual warnings, a push-button alarm test, and a controller battery backup system. The freezer's construction is robust, featuring scratch-resistant cold-rolled steel interior and exterior finishes, and high-density, 2-inch-thick CFC-free urethane foam insulation, which guards against ambient temperature fluctuations. The CFC-free, HFC-blend refrigerant is environmentally friendly. The unit stands on four casters for easy mobility, with the front two being lockable.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Revco ULT3030A20 -30C Lab Freezer includes:
- > Key
- > 4 x Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Revco ULT3030A20 -30C Lab Freezer Specifications
| Temperature Range | -30°C Preset |
| Volume | 29.2 cu. ft. (826L) |
| Refrigerant | R134A & R404A |
| Defrost Method | Automatic |
| Internal Dimensions (H x D x W) | 58 x 29 x 30 in. (147.3 x 73.7 x 76.2 cm) |
| External Dimensions (H x D x W) | 79.2 x 37.2 x 34 in. (201.2 x 94.5 x 86.4 cm) |
| Weight | 445 lb. (197 kg.) |
| Electrical Requirements | 115V, 60Hz, 16A |
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for pixel issues and replace if needed | Completed | |
| Test button functionality and replace any defective buttons | Completed | |
| Examine door seal and lock functionality; replace if compromised | Completed | |
| Conduct PCB test point diagnostics and replace as needed | Completed | |
| Verify compressor and internal fan functionality | Completed | |
| Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
| Inspect all wiring for frays, tears, or breakages | Completed | |
| Test start capacitor capacitance range; replace if needed | Completed | |
| Replace all internal filters and clean condenser fins | Completed | |
| Test alarm battery for full functionality | Completed | |
| Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
| Conduct cosmetic inspection and address any issues | Completed |