Thermo Scientific TSX2330FA High-Performance -30C Lab Freezer
Temperature Range
-35°C to -15°C (-30°C factory preset)
Volume
23.3 cu. ft. (659L)
Defrost Method
Automatic
- V-drive technology adapts compressor speed for energy efficiency and temperature stability
- Forced-air circulation ensures uniform conditions and rapid recovery after door openings
- Microprocessor control system provides precise monitoring and adjustment
- Integrated alarms and setpoint security enhance sample protection
- Standard door locks provide added security for stored materials
- Ergonomic door handle with soft-touch grip supports comfortable use
- Angled control panel reduces strain during operation
- Adjustable epoxy-coated shelves allow flexible storage configurations
- Quiet operation supports a comfortable laboratory environment

An intuitive microprocessor control system allows precise monitoring and adjustment of operating conditions, supported by integrated alarms, setpoint security, and an interior light for improved visibility. The unit features a non-invasive, fully adaptive defrost system that minimizes disruption to stored samples, along with standard door locks for added security. Designed with user comfort in mind, the freezer includes an ergonomically designed door handle with a soft-touch grip, an angled controller to reduce strain during operation, and adjustable epoxy-coated shelves for flexible storage. With quiet operation and a durable design, the Thermo Scientific TSX2330FA High-Performance -30C Lab Freezer delivers dependable performance for laboratory and clinical storage.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific TSX2330FA High-Performance -30C Lab Freezer includes:
- > Key
- > 4 x Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific TSX2330FA High-Performance -30C Lab Freezer Specifications
| Temperature Range | -35°C to -15°C (-30°C factory preset) |
| Volume | 23.3 cu. ft. (659L) |
| Refrigerant | R290 |
| Defrost Method | Automatic |
| Internal Dimensions (W x D x H) | 24 x 28.5 x 58 in. (61 x 72.3 x 147.3 cm) |
| External Dimensions (W x D x H) | 28 x 37.9 x 78.5 in. (71.1 x 96.2 x 199.4 cm) |
| Weight | 371 lb. (168 kg.) |
| Electrical Requirements | 115V, 60Hz, 10A, 1 Ph |
- Recertification Inspection Protocol
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for pixel issues and replace if needed | Completed | |
| Test button functionality and replace any defective buttons | Completed | |
| Examine door seal and lock functionality; replace if compromised | Completed | |
| Conduct PCB test point diagnostics and replace as needed | Completed | |
| Verify compressor and internal fan functionality | Completed | |
| Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
| Inspect all wiring for frays, tears, or breakages | Completed | |
| Test start capacitor capacitance range; replace if needed | Completed | |
| Replace all internal filters and clean condenser fins | Completed | |
| Test alarm battery for full functionality | Completed | |
| Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
| Conduct cosmetic inspection and address any issues | Completed |