Thermo Scientific TSX3020FA High-Performance -20C Lab Freezer
Temperature Range
-25°C to -15°C (-20°C factory preset)
Volume
29.2 cu. ft. (827L)
Defrost Method
Manual
- V-drive compressor technology adapts speed based on usage patterns for efficient operation
- Cold-wall convection cooling supports uniform and stable temperature distribution
- Microprocessor control panel enables precise temperature monitoring and management
- Integrated alarms notify users of critical system conditions
- Interior chamber lighting improves visibility when accessing stored materials
- Standard door lock provides added protection for valuable samples
- Self-closing door with 90 degree stop simplifies loading and inventory handling
Equipped with an intuitive microprocessor control system, the TSX3020FA provides precise temperature management along with setpoint security, integrated alarms, and an interior light for clear visibility when accessing stored materials. Standard door locks add an extra level of protection, while the self-closing door with a 90 degree stop simplifies loading and inventory organization. The Thermo Scientific TSX3020FA High-Performance -20C Lab Freezer enables reliable temperature control, energy-efficient operation, and user-friendly operation to support consistent laboratory storage performance.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific TSX3020FA High-Performance -20C Lab Freezer includes:
- > Key
- > 4 x Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific TSX3020FA High-Performance -20C Lab Freezer Specifications
| Temperature Range | -25°C to -15°C (-20°C factory preset) |
| Volume | 29.2 cu. ft. (827L) |
| Refrigerant | R290 |
| Defrost Method | Manual |
| Internal Dimensions (H x D x W) | 30 x 28.5 x 58 in. (76.2 x 72.3 x 147.3 cm) |
| External Dimensions (H x D x W) | 34 x 38.9 x 78.5 in. (86.4 x 99 x 199.4 cm) |
| Weight | 483 lb. (219 kg.) |
| Electrical Requirements | 115V, 60Hz, 4.7A, 1Ph |
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for pixel issues and replace if needed | Completed | |
| Test button functionality and replace any defective buttons | Completed | |
| Examine door seal and lock functionality; replace if compromised | Completed | |
| Conduct PCB test point diagnostics and replace as needed | Completed | |
| Verify compressor and internal fan functionality | Completed | |
| Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
| Inspect all wiring for frays, tears, or breakages | Completed | |
| Test start capacitor capacitance range; replace if needed | Completed | |
| Replace all internal filters and clean condenser fins | Completed | |
| Test alarm battery for full functionality | Completed | |
| Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
| Conduct cosmetic inspection and address any issues | Completed |

Click Here to View the Technical Specs for the Thermo Scientific TSX3020FA High-Performance -20C Lab Freezer