Tuttnauer 3850EP Digital Autoclave, 230V
The Tuttnauer 3850EP Digital Autoclave 230V, is designed for seamless integration into work environments and features a compact design with heat insulation for efficiency and safety. Constructed with a durable 316L type stainless steel chamber and door, enhanced with an electro-polish finish, this autoclave ensures reliability and longevity. It is equipped with a control lock-out switch to prevent cycle initiation if the door is not properly locked, alongside door failure protection and automatic safety shutoff mechanisms to safeguard against high pressure and overheating, respectively. This model boasts rapid cycles for increased availability of sterile instruments, with a user-friendly, single-touch control system that enhances productivity. This model also includes a built-in printer, which provides a printed summary at the end of each cycle, documenting temperature, elapsed sterilization and drying time, selected cycle number, real time and date, as well as any errors or interruptions that may occur during the cycle.
The 3850EP offers 4 pre-set sterilization programs, customizable cycle parameters for flexibility, and precise control systems for accurate results. With features like interdependent temperature/pressure monitoring and cycle information recovery after power failures, the 3850EP is designed for enhanced monitoring and minimal downtime. Additionally, it includes an insufficient water detector with automatic shutoff for added safety. With an 65L (17.2 gal.) chamber volume and sterilization temperature range of 121°C to 134°C, this autoclave ensures efficient sterilization, boasting standard unwrapped cycle times of 31 minutes cold and 20 minutes hot, making it a reliable choice for demanding sterilization needs. Please Note: This Autoclave operates at 230V ONLY and is not capable of autoclaving liquids.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Tuttnauer 3850EP Digital Autoclave, 230V includes:
- > 2 x Stainless-Steel Trays
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts & Labor Warranty
Tuttnauer 3850EP Digital Autoclave, 230V Specifications
| Chamber Volume | 65L (17.2 gal.) |
| Sterilization Temperature Range | 121°C to 134°C |
| Standard Unwrapped Cycle Time |
Cold: 31 minutes Hot: 20 minutes |
| Standard Cassettes Capacity |
10 full |
| Maximum Available Height |
12.1 in. (30.7 cm) |
| Number of Trays |
2 |
| Tray Dimensions (W x D x H) |
11 x 20 x 1 in. (28 x 50.8 x 2.5 cm) 14 x 20 x 1 in. (35.5 x 50.8 x 2.5 cm) |
| Chamber Dimensions (Diameter x D) | 15 x 20 in. (38 x 50.8 cm) |
| Overall Dimensions (W x H x D) | 26 x 21 x 27 in. (66 x 53.3 x 68.6 cm) |
| Weight | 196 lb. (89 kg) |
| Electrical Requirements | 230V, 50/60Hz, 10.4A, 2 Ph |
- Verify power supply output to ensure precise voltage delivery as per equipment specifications
- Check full display functionality, replacing any dead or dim pixels for optimal clarity
- Confirm unit is level to maintain stability and accurate operational performance
- Inspect door seal integrity, replacing if any wear or damage compromises airtight performance
- Examine and clean both external and internal filters to ensure unobstructed airflow and system efficiency
- Execute comprehensive cycle testing to validate consistent operation
- Verify printer functionality (if applicable) to ensure precise print outputs without error
- Assess internal plumbing system for any leaks or pressure inconsistencies
- Inspect all electrical connections and wiring for secure attachment and optimal conductivity
- Verify functionality of internal switches to ensure reliable engagement and operational control
- Inspect and clean the water reservoir to maintain purity and prevent contamination
- Confirm accuracy of pressure readings, recalibrating if deviations are identified
- Perform final test cycles, ensuring zero errors for validated operational readiness
Technician Inspection and Recertification Checklist: Autoclaves
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltages | Completed | |
| Check display functionality and replace dead/dim pixels | Completed | |
| Confirm unit leveling for stability | Completed | |
| Inspect and replace damaged door seal | Completed | |
| Clean external and internal filters for airflow efficiency | Completed | |
| Conduct comprehensive cycle testing | Completed | |
| Confirm printer functionality (if applicable) | Completed | |
| Assess plumbing for leaks or pressure issues | Completed | |
| Inspect electrical connections for secure and optimal conductivity | Completed | |
| Verify internal switch functionality | Completed | |
| Inspect and clean water reservoir | Completed | |
| Confirm pressure reading accuracy and recalibrate if needed | Completed | |
| Perform final test cycles with zero errors | Completed |