Tuttnauer 3870M Manual Autoclave, 230V
Temperature Range
100°C to 134°C
Chamber Volume
85L (22 gal.)
Chamber Dimensions (Dia. x L)
15 x 27 in. (38.1 x 68.6 cm)
- Automatic shut-off at the end of sterilization and drying cycles for enhanced energy efficiency
- Double locking safety device prevents door opening while the chamber is pressurized
- Durable electro-polished chamber and door construction ensure long-lasting performance
- Accommodates various sterilization needs, including glassware, wrapped, and unwrapped instruments
- Combination pressure and temperature gauge allows easy monitoring during operation
- Simple front-located valve enables convenient reservoir drainage
- Includes a three-shelf stand for efficient storage and organization
- Fast cycle times with 23-minute hot starts and 32-minute cold starts for quick sterilization
Tuttnauer 3870M Manual Autoclave, 230V is equipped for high-volume use, suitable for sterilizing both long instruments and numerous small tools. It operates with an automatic shut-off at the end of sterilization and drying cycles for energy efficiency. Safety features include a double locking device to prevent the door from opening under pressure and an automatic shut-off to avoid overheating. It offers easy storage with an accompanying three-shelf stand and simple front-located valve for reservoir drainage.
The chamber and door are constructed for longevity, with an electro-polished finish. The autoclave can handle a range of sterilization needs, accommodating glass, wrapped, and unwrapped tools, with temperature settings from 100°C to 134°C. Cycle times vary, typically taking 23 minutes after a hot start and 32 minutes from a cold start. It also includes a combination pressure and temperature gauge for monitoring. Please Note: This Autoclave operates at 230V ONLY.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Tuttnauer 3870M Manual Autoclave, 230V includes:
- > 2 x Stainless-Steel Trays
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts & Labor Warranty
Tuttnauer 3870M Manual Autoclave, 230V Specifications
| Chamber Volume | 85L (22 gal.) |
| Sterilization Temperature Range | 100°C to 134°C |
| Standard Unwrapped Cycle Time |
Cold: 32 minutes Hot: 23 minutes |
| Standard Cassettes Capacity |
15 |
| Maximum Available Height |
12.1 in. (30.7 cm) |
| Number of Trays |
2 |
| Tray Dimensions |
26 x 11 x 1 in. (67 x 28 x 2.5 cm) 26 x 14 x 1 in. (67 x 35 x 2.5 cm) |
| Chamber Dimensions (Diameter x L) | 15 x 27 in. (38.1 x 68.6 cm) |
| Overall Dimensions (W x H x D) | 26 x 21 x 34 in. (66 x 53.3 x 86.4 cm) |
| Weight | 278 lb. (126 kg) |
| Electrical Requirements | 230V, 60Hz, 1 Phase |
- Verify power supply output to ensure precise voltage delivery as per equipment specifications
- Check full display functionality, replacing any dead or dim pixels for optimal clarity
- Confirm unit is level to maintain stability and accurate operational performance
- Inspect door seal integrity, replacing if any wear or damage compromises airtight performance
- Examine and clean both external and internal filters to ensure unobstructed airflow and system efficiency
- Execute comprehensive cycle testing to validate consistent operation
- Verify printer functionality (if applicable) to ensure precise print outputs without error
- Assess internal plumbing system for any leaks or pressure inconsistencies
- Inspect all electrical connections and wiring for secure attachment and optimal conductivity
- Verify functionality of internal switches to ensure reliable engagement and operational control
- Inspect and clean the water reservoir to maintain purity and prevent contamination
- Confirm accuracy of pressure readings, recalibrating if deviations are identified
- Perform final test cycles, ensuring zero errors for validated operational readiness
Technician Inspection and Recertification Checklist: Autoclaves
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltages | Completed | |
| Check display functionality and replace dead/dim pixels | Completed | |
| Confirm unit leveling for stability | Completed | |
| Inspect and replace damaged door seal | Completed | |
| Clean external and internal filters for airflow efficiency | Completed | |
| Conduct comprehensive cycle testing | Completed | |
| Confirm printer functionality (if applicable) | Completed | |
| Assess plumbing for leaks or pressure issues | Completed | |
| Inspect electrical connections for secure and optimal conductivity | Completed | |
| Verify internal switch functionality | Completed | |
| Inspect and clean water reservoir | Completed | |
| Confirm pressure reading accuracy and recalibrate if needed | Completed | |
| Perform final test cycles with zero errors | Completed |