VWR Digital 6 Block Heater, cat # 12621-100
The VWR Digital 6 Place Heat Block accommodates 6 blocks and features high wattage, along with constant-temperature heaters that are economical, versatile, and compact. This model accepts several interchangeable blocks, accommodating various tubes sizes from 0.25ml micro to 50ml. Each block has a thermometer well. The efficient heater features close tube-and-block contact with a low density design for exceptional temperature uniformity. Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire.
All of our equipment is fully tested before being shipped.
The VWR Digital 6 Block Heater, cat # 12621-100 includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Digital 6 Block Heater, cat # 12621-100 Specifications
|
Temperature Range |
Ambient +5C to 120C |
|
Temperature Stability |
0.1C |
|
Temperature Uniformity |
0.3C |
|
Heat Up Time to 100C |
75 minutes |
|
Dimensions |
20.9 x 8 x 3.5 in. (53.1 x 20.3 x 8.9
cm)
|
|
Weight |
10 lbs. (4.5 kg) |
|
Electrical Requirements |
120V, 50/60Hz, 3.4 amps, 410W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |
VWR Digital Heat Block Manual
Click Here to View all the Technical Specs for the VWR 6 Block Heater