VWR Digital Advanced 6 Block Heater Cat. # 75838-306
The VWR Digital Advanced 6 Block Heater, Cat. # 75838-306, accommodates up to six blocks and features high wattage, constant-temperature heaters that are economical, versatile, and compact. This model supports several interchangeable blocks to accommodate a wide range of tube sizes from 0.25 mL micro to 50 mL. Each block includes a thermometer well for precise monitoring. The heater’s efficient design ensures close tube-and-block contact with exceptional temperature uniformity, making it ideal for applications such as incubation, enzyme reactions, immunoassays, and other laboratory procedures. Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire.
All of our equipment is fully tested before being shipped.
The VWR Digital Advanced 6 Block Heater Cat. # 75838-306 includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Digital Advanced 6 Block Heater Cat. # 75838-306 Specifications
| Temperature Range | Ambient +5°C to 120°C |
| Temperature Stability | ± 0.1°C at 37°C |
| Temperature Uniformity |
Within block: ± 0.1°C at 37°C Across similar blocks ± 0.3°C at 37°C |
| Heat Up Time to 100°C | 75 minutes |
| Dimensions (L x W x H) | 20.7 x 7.9 x 4.1 in (52.6 x 20.1 x 10.4 cm) |
| Weight | 10 lb. (4.5 kg) |
| Electrical Requirements | 120 V, 50/60 Hz, 410W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |