VWR Digital Heat Block, cat# 12621-084
Temperature Range
Ambient +5°C to 120°C
Temperature Uniformity
± 0.1°C at 37°C
Block Capacity
1 Heat Block
- Compact and versatile dual block heater design
- High wattage heating elements for efficient performance
- Accepts interchangeable blocks for various tube sizes
- Close tube-and-block contact ensures optimal heat transfer
- Durable construction for long-term laboratory use
- Compatible with multiple block configurations
- Easy-to-use digital interface for precise control
The VWR Digital Heat Block, Cat. #12621-084, 100 Watt, multi-purpose unit is adjustable with high and low thermostats that provide reproducible and accurate temperature control with a stability of ± 2°C. Temperature range is ambient to 150C. Please Note: A Heat block is NOT INCLUDED with this Dry Bath Heater. If you require a heat block, please inquire.
All of our equipment is fully tested before being shipped.
The VWR Analog Dual Heat Block, cat # 12621-108 includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
- > Please Note: A heat block is NOT INCLUDED with this Dry Bath Heater. If you require a heat block, please inquire
VWR Digital Heat Block, cat# 12621-084 Specifications
|
Temperature Range |
Ambient +5°C to 120°C |
|
Temperature Stability |
± 0.1°C |
|
Temperature Uniformity |
± 0.1C |
|
Dimensions (L x W x H) |
12.4 x 8 x 3.5 in (31.5 x 20.3 x 8.9 cm) |
|
Electrical Requirements |
120V, 50/60Hz, 0.9A, 110W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |


Click Here to View the Technical Specs for the VWR Heat Block