Temperature Range
Ambient +5C to 120C
Temperature Uniformity
±
0.1C at 37C
Block Capacity
4 Heat Blocks
- Compact and versatile dual block heater design
- High wattage heating elements for efficient performance
- Accepts interchangeable blocks for various tube sizes
- Close tube-and-block contact ensures optimal heat transfer
- Durable construction for long-term laboratory use
- Easy-to-use digital interface for precise control
All of our equipment is fully tested before being shipped.
The VWR Digital Multi Block Heater, Cat. #: 10027-044 includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Digital Dual Heat Block Specifications
| Temperature Range | Ambient +5°C to 120°C |
| Temperature Stability | 0.1°C at 37°C |
| Temperature Uniformity | ± 0.1°C at 37°C |
| Heat Up Time to 100C | 70 minutes |
| Dimensions | 8 x 16.9 x 3.5 in. (20.3 x 42.9 x 8.9 cm) |
| Weight | 8.5 lbs. (3.9 kg) |
| Electrical Requirements | 120V, 50/60Hz, 2.6A, 310W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |