Temperature Uniformity
0.1C
Temperature Range
Ambient +5C to 120C
Dimensions
18.4 x 8 x 3.5 in (46.7 x 20.3 x 8.9 cm)
- Accommodates 3 blocks for multiple uses.
- High wattage, constant-temperature heaters for efficient and compact heating.
- Compatible with a range of tube sizes from 0.25mL micro to 50mL.
- Efficient heater design ensures close contact between tube and block, providing uniform temperature.
The VWR digital triple heat block accomodates 3 blocks and features high wattage, constant-temperature heaters that are economical, versatile, and compact. This model accepts several interchangeable blocks, accommodating various tube sizes from 0.25mL micro to 50mL. Efficient heater features close tube-and-block contact with a low density design for exceptional temperature uniformity. Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire.
All of our equipment is fully tested before being shipped.
The VWR Digital Triple Heat Block, cat# 12621-092 includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Digital Triple Heat Block Specifications
|
Temperature Range |
Ambient +5C to 120C |
|
Temperature Stability |
0.1C |
|
Temperature Uniformity |
0.1C |
|
Dimensions (L x W x H) |
18.4 x 8 x 3.5 in (46.7 x 20.3 x 8.9 cm) |
|
Electrical Requirements |
120V, 50/60Hz, 1.7A, 210W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |