VWR Mega Star 1.6R Refrigerated Centrifuge
Maximum Speed
15,200 RPM (25,830 x g RCF)
Maximum Capacity
4 x 400mL
Temperature Range
-10°C to 40°C
- Refrigerated design enhances sample security and reproducibility across workflows
- Intuitive keypad controls provide easy setup and clear access to operating parameters
- Accessible rotor chamber allows for fast cleaning and improved laboratory hygiene
- Compact benchtop design maximizes workspace without sacrificing performance
- Ergonomic height and curved design simplify sample loading and unloading
- Optimized for tissue culture and general-purpose separation workflows
- Energy-efficient operation reduces power consumption during standard protocols
- Designed to streamline workflow transitions and minimize interruptions

Engineered with the Auto-Lock rotor system, the Mega Star 1.6R enables quick, simple rotor installation and exchange, reducing downtime between applications. The accessible rotor chamber supports fast cleaning and promotes a healthier working environment. With configurations tailored for a variety of sample vessels, the system is well-suited for a number of standard research applications.
Its ergonomic, space-saving design optimizes bench space without limiting functionality. The curved profile and accessible height make loading and unloading samples straightforward, while also simplifying routine maintenance. In addition, the system is energy efficient, offering significant reductions in energy consumption during common protocols. Altogether, the Mega Star 1.6R delivers a balance of performance, convenience, and efficiency for modern laboratory workflows.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The VWR Mega Star 1.6R Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Guide
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
VWR Mega Star 1.6R Refrigerated Centrifuge Specifications
| Maximuim Speed | 15,200 RPM |
| Minimum Speed | 300 RPM |
| Maximum RCF | 25,830 x g |
| Maximum Capacity | 4 x 400ml |
| Temperature Range | -10°C to 40°C |
| Timer | Up to 9 hours 59 minutes, and continuous run |
| Accleration Rates | 9 |
| Deceleration Rates | 10 |
| Noise Level at Max Speed | < 68 dB (A) |
| Dimensions (W x D x H) | 24.6 x 33.6 x 14.2 in. (62.5 x 85.5 x 36.1 cm) |
| Weight | 202.8 lbs. (92 kg) |
| Electrical Requirements | 120V, 60Hz, 11A, 1350W |
Click Here to Expand Rotor Specifications
|
TX-200 Rotor Specs |
|
|
Capacity |
4 x 180 mL bottles |
|
Max RPM |
5,500 |
|
Max RCF |
5,580 x g |
|
Fiberlite F-15-6x100Y Rotor Specs |
|
|
Capacity |
6 x 100 mL Tubes |
|
Max RPM |
13,000 |
|
Max RCF |
18,516 x g |
|
M-20 Microplate Rotor Specs |
|
| Capacity |
2 x 3 Standard or 2 x 1 Deepwell Plates |
|
Max RPM |
4,000 |
|
Max RCF |
2,272 x g |
|
TX-400 Rotor Specs |
|
| Capacity |
4 x 400 mL Bottles |
|
Max RPM |
4,700 RPM |
|
Max RCF |
4,149 x g |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |


Click Here to View the Technical Specs for the VWR Mega Star 1.6R Refrigerated Centrifuge