Wescor Cytopro 7620 Cytocentrifuge
Speed Range
100 to 2,000 RPM
Capacity
Up to 8 Slides
Sample Chamber Sizes
Double: 0.3ml per chamber
Single: 0.5ml per chamber
- Eight-station rotor for processing multiple samples efficiently
- Patented Cytopro system with touch pad controls for ease of use
- Dual sample chambers enable increased efficiency and economy
- Pre-wetting fluid ports for in situ fixation and enhanced versatility
- Supports 9 programmable cycles for flexible workflow management
- 4-line by 16-character LCD screen for clear, detailed display
- Compact design with dimensions of 17 x 9.5 x 21 in. for space-saving setup
- Lightweight construction at 27 lb. for easy portability
- Compatible with 100-120 or 220-240V, 50/60Hz electrical configurations
As the process begins, the cell suspension is pipetted into the sample well or the tunnel. As the rotor starts spinning, centrifugal force moves the suspension towards the slide. Cells sediment onto the slide as the Cytopad absorbs the suspension medium. Cytopads come in two absorption speeds that are color-coded for ease of use (not included). The white pad is for faster absorption of thick debris-filled suspension fluids. The brown pad is for slower absorption of thin fluids such as cerebral spinal fluid (CSF) where slower flow enhances cell recovery. The Wescor Cytopro 7620 Cytocentrifuge puts unsurpassed ease of use, maximum flexibility, and safety at the user’s command.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Wescor Cytopro 7620 Cytocentrifuge includes:
- > Cytopro Rotor
- > 8 Cytoclips
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Wescor Cytopro 7620 Cytocentrifuge Specifications
| Speed Range | 100 to 2,000 RPM |
| Capacity | Up to 8 Slides |
| Sample Chamber Sizes |
Double: 0.3ml per chamber Single: 0.5ml per chamber |
| Programs | 9 |
| Timer | 1 to 99 minutes |
| Display | 4 line by 16 character LCD screen |
| Overall Dimensions (W x H x D) | 17 x 9.5 x 21 in. (43 x 24 x 53 cm) |
| Weight | 27 lb. (12.25 kg.) |
| Electrical Requirements | 100-120 or 220-240V, 50/60Hz, 75W |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |



Click Here to View the Technical Specs for the Wescor Cytopro 7620 Cytocentrifuge